Suppr超能文献

一种基于符合方法的多普勒微栓子信号识别新技术:体外和体内评估

A novel technique for identification of doppler microembolic signals based on the coincidence method: in vitro and in vivo evaluation.

作者信息

Georgiadis D, Goeke J, Hill M, König M, Nabavi D G, Stögbauer F, Zunker P, Ringelstein E B

机构信息

Department of Neurology, University of Münster, Germany.

出版信息

Stroke. 1996 Apr;27(4):683-6. doi: 10.1161/01.str.27.4.683.

Abstract

BACKGROUND AND PURPOSE

The applicability of a novel differentiation technique in embolus detection based on the coincidence principle and using a multigate probe was evaluated in this study.

METHODS

According to the coincidence method, high-intensity transients should only be classified as microembolic signals if they appear sequentially in the two sample volumes monitored and within a defined time window calculated from the blood velocity and the spatial distance between the insonation depths. Part A: microbubbles were introduced in a continuous flow bench model of the middle cerebral artery to evaluate the accuracy of the multigate probe in embolus detection. Part B: in the subjects and patients, the minimal and maximum time delays in the appearance of microembolic signals in the two middle cerebral artery sample volumes were calculated as 0.01 second and set at 0.1 second, respectively. The multigate probe was used to monitor (1) 5 normal volunteers in whom 1008 artifact signals were produced,(2) 2 patients undergoing aortic valve replacement surgery, and (3) 12 patients with potential cardiac or carotid embolic sources.

RESULTS

In the bench model, 95.5% of microembolic signals produced by microbubbles appeared in the two sample volumes with a time delay between 0.02 and 0.05 second, while in the remaining 4.5% a shorter passage time of 0.01 second was measured. A total of 1968 high-intensity signals were recorded in subjects and patients. All but 20 of these (99%) appeared in both monitoring channels within the above time frame. To summarize, 996 (98.8%) of the 1004 artifact signals and 943 (98.1%) of the 961 microembolic signals were correctly classified.

CONCLUSIONS

Application of the coincidence theory to distinguish microembolic signals from artifacts provides a promising new technique with high sensitivity and specificity that could decisively improve the validity of embolus detection.

摘要

背景与目的

本研究评估了一种基于符合原理并使用多门控探头的新型鉴别技术在栓子检测中的适用性。

方法

根据符合方法,只有当高强度瞬态信号在两个监测样本容积中按顺序出现,且在根据血流速度和声束深度之间的空间距离计算出的定义时间窗内时,才应将其分类为微栓子信号。A部分:在大脑中动脉连续血流实验台模型中引入微泡,以评估多门控探头在栓子检测中的准确性。B部分:在受试者和患者中,计算大脑中动脉两个样本容积中微栓子信号出现的最小和最大时间延迟分别为0.01秒和0.1秒。使用多门控探头监测:(1)5名产生1008个伪像信号的正常志愿者;(2)2名接受主动脉瓣置换手术的患者;(3)12名有潜在心脏或颈动脉栓子来源的患者。

结果

在实验台模型中,微泡产生的微栓子信号中有95.5%出现在两个样本容积中,时间延迟在0.02至0.05秒之间,而其余4.5%的通过时间较短,为0.01秒。在受试者和患者中共记录到1968个高强度信号。其中除20个(99%)外,所有信号均在上述时间框架内在两个监测通道中出现。总之,1004个伪像信号中的996个(98.8%)和961个微栓子信号中的943个(98.1%)被正确分类。

结论

应用符合理论区分微栓子信号和伪像提供了一种有前景的新技术,具有高灵敏度和特异性,可显著提高栓子检测的有效性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验