Parkes R B, Kolstad R A
J Periodontol. 1982 Jul;53(7):434-8. doi: 10.1902/jop.1982.53.7.434.
The sharpness of surgical instruments is of particular importance to the periodontist. In order to study the effects of sterilization on the cutting edges of periodontal curets and scalers, new sharpened instruments were sterilized 10 cycles with saturated steam, formalin-alcohol vapor or dry heat. The instruments were evaluated before and after sterilization with a scanning electron microscope. No visual damage to the cutting edges of stainless steel instruments was evident after sterilization by saturated steam at 250 degrees F, formalin-alcohol vapor at 270 degrees F or dry heat up to 340 degrees F. Carbon steel instruments were not damaged by formalin-alcohol vapor at 270 degrees F or dry heat up to 340 degrees F. Saturated steam caused oxidation and dulling of carbon steel instruments.
手术器械的锋利度对牙周病医生尤为重要。为了研究灭菌对牙周刮治器和洁治器刀刃的影响,将新磨锐的器械用饱和蒸汽、甲醛酒精蒸汽或干热进行10次循环灭菌。在灭菌前后用扫描电子显微镜对器械进行评估。在250华氏度的饱和蒸汽、270华氏度的甲醛酒精蒸汽或高达340华氏度的干热灭菌后,不锈钢器械的刀刃没有明显的视觉损伤。碳钢器械在270华氏度的甲醛酒精蒸汽或高达340华氏度的干热下没有受损。饱和蒸汽会导致碳钢器械氧化和钝化。