Molloi S, Ersahin A, Tang J, Hicks J, Leung C Y
Department of Radiological Sciences, Medical Sciences I, University of California, Irvine, 92717, USA.
Circulation. 1996 May 15;93(10):1919-27. doi: 10.1161/01.cir.93.10.1919.
As a solution to the well-documented problems associated with visual interpretation of coronary arteriograms, more physiological methods of assessing coronary artery stenosis are being investigated. Volumetric coronary blood flow (BF) can be a valuable aid in the analysis of functional significance of arterial obstruction.
The left anterior descending coronary artery (LAD) of 15 anesthetized pigs (40 to 50 kg) was dissected free from the epicardium in its proximal portion, and a transmit-time ultrasound flow probe of the appropriate size was applied. A vascular occluder was positioned distal to the flow probe for flow adjustments. Contrast injections (2 to 4 mL/s for 3 seconds) were made into the left main coronary artery during image acquisition with a motion-immune dual-energy digital subtraction angiography (DE DSA) system. Tissue-suppressed energy-subtracted images were used to generate time-density curves. BF measurements were made in the LAD vascular bed with use of the time-density curve, with consideration that blood was momentarily replaced with contrast during the injection. In 19 comparisons, the mean BF, measured with the use of DE DSA, correlated extremely well with the mean ultrasound flow (DE DSA=0.90 ultrasound+3.10 mL/min, r=.96). Also, contrast injection increased the BF by an average of only 15% during the image-acquisition time interval.
Accurate BF measurements can be made with motion-immune DE DSA. The BF measurements can be completed before the onset of significant changes in BF due to contrast injection. Furthermore, it is possible to make the BF measurements during routine coronary arteriography.
作为解决与冠状动脉造影视觉解读相关的诸多问题的一种方法,更多评估冠状动脉狭窄的生理学方法正在研究中。容积性冠状动脉血流量(BF)在分析动脉阻塞的功能意义方面可能是一项有价值的辅助手段。
对15只麻醉的猪(体重40至50千克)的左前降支冠状动脉(LAD)在其近端从心外膜游离出来,并应用合适尺寸的传输时间超声血流探头。在血流探头远端放置一个血管封堵器以进行血流调节。在使用运动免疫双能数字减影血管造影(DE DSA)系统进行图像采集期间,以2至4毫升/秒的速度向左主干冠状动脉内注射造影剂3秒。使用组织抑制能量减影图像生成时间 - 密度曲线。考虑到注射期间血液被造影剂暂时替代,利用时间 - 密度曲线在LAD血管床进行BF测量。在19次比较中,使用DE DSA测量的平均BF与平均超声血流相关性非常好(DE DSA = 0.90×超声血流 + 3.10毫升/分钟,r = 0.96)。此外,在图像采集时间间隔内,造影剂注射使BF平均仅增加15%。
使用运动免疫DE DSA可进行准确的BF测量。BF测量可在因造影剂注射导致BF发生显著变化之前完成。此外,在常规冠状动脉造影期间进行BF测量是可行的。