Petroll W M, Jester J V, Cavanagh H D
Department of Opthalmology, University of Texas Southwestern Medical Center, Dallas 75235-9057, USA.
Scanning. 1996 Jan;18(1):45-9. doi: 10.1002/sca.1996.4950180107.
A new depth encoding system (DES) is presented, which makes it possible to calculate, display, and record the z-axis position continuously during in vivo imaging using tandem scanning confocal microscopy (TSCM). In order to verify the accuracy of the DES for calculating the position of the focal plane in the cornea both in vitro and in vivo, we compared TSCM measurements of corneal thickness to measurements made using an ultrasonic pachymeter (UP, a standard clinical instrument) in both enucleated rabbit, cat, and human eyes (n = 15), and in both human patients (n = 7). Very close agreement was found between the UP and TSCM measurements in enucleated eyes; the mean percent difference was 0.50 +/- 2.58% (mean +/- SD, not significant). A significant correlation (R = 0.995, n = 15, p < 0.01) was found between UP and TSCM measurements. These results verify that the theoretical equation for calculating focal depth provided by the TSCM manufacturer is accurate for corneal imaging. Similarly, close agreement was found between the in vivo UP and TSCM measurements; the mean percent differences was 1.67 +/- 1.38% (not significant), confirming that z-axis drift can be minimized with proper applanation of the objective. These results confirm the accuracy of the DES for imaging of the cornea both ex vivo and in vivo. This system should be of great utility for applications where quantitation of the three-dimensional location of cellular structures is needed.
本文介绍了一种新的深度编码系统(DES),该系统能够在使用串联扫描共聚焦显微镜(TSCM)进行体内成像时连续计算、显示和记录z轴位置。为了验证DES在体外和体内计算角膜焦平面位置的准确性,我们将TSCM测量的角膜厚度与使用超声角膜测厚仪(UP,一种标准临床仪器)在摘除的兔眼、猫眼和人眼(n = 15)以及人类患者(n = 7)中进行的测量进行了比较。在摘除的眼中,UP和TSCM测量结果非常接近;平均百分比差异为0.50 +/- 2.58%(平均值 +/- 标准差,无显著性差异)。UP和TSCM测量之间发现了显著相关性(R = 0.995,n = 15,p < 0.01)。这些结果验证了TSCM制造商提供的计算焦深的理论方程对于角膜成像来说是准确的。同样,在体内UP和TSCM测量之间也发现了密切的一致性;平均百分比差异为1.67 +/- 1.38%(无显著性差异),证实了通过适当压平物镜可以将z轴漂移降至最低。这些结果证实了DES在体外和体内角膜成像中的准确性。该系统对于需要定量细胞结构三维位置的应用应该非常有用。