氙/CT与凯蒂-施密特(氧化亚氮)法所获脑血流量值之间的相关性。

Correlation between cerebral blood flow values obtained by Xenon/CT and Kety-Schmidt (N2O) methods.

作者信息

Tone O, Ito U, Tomita H, Akimoto H, Sakemi H

机构信息

Department of Neurosurgery, Musashino Red Cross Hospital, Tokyo Japan.

出版信息

Acta Neurol Scand Suppl. 1996;166:18-21. doi: 10.1111/j.1600-0404.1996.tb00535.x.

Abstract

The means of the cerebral blood flow (CBF) values obtained by the stable xenon enhanced CT (Xe/CT) method using two different CT scanners were compared with global CBF value obtained by the Kety-Schmidt (N2O) method as a reference. Xe/CT CBF values were obtained using a GE CT9200 (31 patients, 2 flow maps, 120 kV, absorption constant of 0.040) as well as a GE ProSeed Accell (38 patients, 4 flow maps, 80 kV, absorption constant of 0.028). The protocol of inhalation in the Xe/XT method consisted of 4 min wash-in and 4-min wash-out of 35% stable xenon. In the Kety-Schmidt method, 15% N2O gas was inhaled for 10 min. The N2O content of blood samples was measured using a van Slyke-Neill blood gas analyzer. We corrected all obtained CBF values for a PaCO2 of 34 mmHG (CBF34). The global CBF34 values obtained by the Kety-Schmidt method were linearly correlated with the CBF34 values obtained using the CT9200 and with those obtained using the ProSeed Accell, and the regression line equations were, respectively, Y = 0.64X + 13.7 (X: CT9200, Y: Kety-Schmidt, r = 0.666, p < 0.01) and Y = 0.99X + 11.2 (X: ProSeed Accell, Y: Kety Schmidt, r = 0.756, p < 0.01). Since the CBF values obtained by the Xe/CT method using different CT scanners are not always the same as the global CBF values obtained by the Kety-Schmidt method, CBF values obtained by the Xe/CT method should be corrected referring to the regression line obtained by applying both methods for each patient.

摘要

将使用两种不同CT扫描仪的稳定氙增强CT(Xe/CT)方法获得的脑血流量(CBF)值的均值,与作为参考的通过Kety-Schmidt(N2O)方法获得的全脑CBF值进行比较。使用GE CT9200(31例患者,2张血流图,120 kV,吸收常数0.040)以及GE ProSeed Accell(38例患者,4张血流图,80 kV,吸收常数0.028)获得Xe/CT CBF值。Xe/XT方法中的吸入方案包括35%稳定氙的4分钟洗入和4分钟洗出。在Kety-Schmidt方法中,吸入15% N2O气体10分钟。使用范斯莱克-尼尔血气分析仪测量血样中的N2O含量。我们将所有获得的CBF值校正为PaCO2为34 mmHg时的值(CBF34)。通过Kety-Schmidt方法获得的全脑CBF34值与使用CT9200获得的CBF34值以及使用ProSeed Accell获得的CBF34值呈线性相关,回归线方程分别为Y = 0.64X + 13.7(X:CT9200,Y:Kety-Schmidt,r = 0.666,p < 0.01)和Y = 0.99X + 11.2(X:ProSeed Accell,Y:Kety Schmidt,r = 0.756,p < 0.01)。由于使用不同CT扫描仪的Xe/CT方法获得的CBF值并不总是与Kety-Schmidt方法获得的全脑CBF值相同,因此对于通过Xe/CT方法获得的CBF值,应参考针对每位患者应用两种方法所获得的回归线进行校正。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验