A detailed study of dose data has been made for 100 coronary angiography examinations performed on a digital X-ray unit. Dose-area product (DAP) data have been analysed in terms of fluoroscopy and radiography for different projections. Projections patterns were similar for all cardiologists studied and the contribution to DAP from fluoroscopy was 32%. Data for 10 patients, selected because the percentage contributions to DAP from fluoroscopy and radiography were similar to the mean for the group, were used to derive conversion factors from DAP to effective dose. The mean DAP was 14 Gy cm2 and the effective dose 3.1 mSv. Conversion factors, based on the same protocol, were derived for a range of tube potentials and filtrations for use in estimation of effective doses with other units.

摘要

对在数字X射线设备上进行的100例冠状动脉造影检查的剂量数据进行了详细研究。已根据不同投照的透视和摄影对剂量面积积(DAP)数据进行了分析。所有研究的心脏病专家的投照模式相似,透视对DAP的贡献为32%。选择了10例患者的数据,因为其透视和摄影对DAP的百分比贡献与该组的平均值相似,用于推导从DAP到有效剂量的转换因子。平均DAP为14 Gy cm²,有效剂量为3.1 mSv。基于相同方案,针对一系列管电压和滤过情况推导了转换因子,用于估算其他设备的有效剂量。