Long-duration metal-on-metal total hip arthroplasties with low wear of the articulating surfaces.

作者信息

Schmalzried T P, Peters P C, Maurer B T, Bragdon C R, Harris W H

机构信息

Joint Replacement Institute at Orthopaedic Hospital, Los Angeles, CA 90007, USA.

出版信息

J Arthroplasty. 1996 Apr;11(3):322-31. doi: 10.1016/s0883-5403(96)80085-4.

Abstract

The 20-year performance of metal-on-metal hip articulations has not been reported. Five McKee-Farrar total hip prostheses and one Sivash prosthesis were obtained at revision surgery after a mean implantation time of 21.3 years. A radiographic, histologic, implant, and wear analysis was performed on these total hip implants with cobalt-chrome metal-on-metal articulations. All cases were associated with femoral component loosening, but the bearing surfaces performed remarkably well. The worst case estimate of combined femoral and acetabular linear wear was 4.2 microns per year, about 25 times less than that typically seen with polyethylene. Metal particles and foreign-body inflammation were seen in all cases, but the volume of reactive tissue was small compared with what is generally seen at revision of hips with a polyethylene acetabular bearing. This may be due to a reduced particle burden or a decreased inflammatory reaction to particulate metal, or both. In addition to articular wear, other sources of metal particles included femoral neck impingement on the acetabular rim, stem burnishing, and corrosion. Prosthetic hip reconstructions can fail for many reasons, including suboptimal femoral stem and/or acetabular cup design and/or fixation. By today's standards, the McKee-Farrar and Sivash stem and acetabular component designs are suboptimal; however, after more than 20 years of use, the metal-on-metal bearing surfaces in these cases demonstrated low wear and do not appear to be the cause of failure. Recent advances in total hip arthroplasty, which include improved implant design, materials, manufacturing, and fixation, combined with a better understanding of the mechanisms of implant loosening and failure, suggest that the cobalt-chrome metal-on-metal bearing be reexamined as an alternative to polyethylene when exceptional durability is required.

摘要

金属对金属髋关节置换的20年性能尚未见报道。在平均植入21.3年后的翻修手术中获取了5个麦基 - 法拉尔全髋关节假体和1个西瓦什假体。对这些采用钴铬金属对金属关节的全髋关节植入物进行了影像学、组织学、植入物及磨损分析。所有病例均伴有股骨部件松动,但关节面表现良好。股骨和髋臼线性磨损的最坏情况估计为每年4.2微米,约为聚乙烯通常所见磨损的25分之一。所有病例中均可见金属颗粒和异物炎症,但与聚乙烯髋臼承重的髋关节翻修时通常所见相比,反应性组织的体积较小。这可能是由于颗粒负荷降低或对颗粒金属的炎症反应减弱,或两者皆有。除了关节磨损外,金属颗粒的其他来源包括股骨颈对髋臼边缘的撞击、柄部打磨和腐蚀。人工髋关节重建失败可能有多种原因,包括股骨柄和/或髋臼杯设计和/或固定欠佳。按照当今的标准,麦基 - 法拉尔和西瓦什的柄部及髋臼部件设计欠佳;然而,经过20多年的使用,这些病例中的金属对金属关节面显示出低磨损,似乎并非失败的原因。全髋关节置换术的最新进展,包括改进的植入物设计、材料、制造和固定,以及对植入物松动和失败机制的更好理解,表明在需要卓越耐久性时,钴铬金属对金属承重可作为聚乙烯的替代方案重新审视。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验