Measurement of regional aortic compliance by MR imaging: a study of reproducibility.

作者信息

Forbat S M, Mohiaddin R H, Yang G Z, Firmin D N, Underwood S R

机构信息

Royal Brompton Hospital, London, UK.

出版信息

J Magn Reson Imaging. 1995 Nov-Dec;5(6):635-9. doi: 10.1002/jmri.1880050604.

Abstract

The reproducibility of MR imaging for the measurement of aortic compliance was studied in 47 healthy volunteers. Long and short term reproducibility and intraobserver variability were tested. The method was modified to improve image quality and short term reproducibility and intraobserver variability retested. For comparison, spin echo imaging was compared with cine gradient echo imaging. Initial long term reproducibility showed a mean difference (+/-SE) of 3% (+/- 7%) with 95% confidence interval (CI) for limits of agreement of +/- 69%. Short term reproducibility (7% +/- 6%, 95% CI +/- 46%) and intraobserver variability (1% +/- 2%, 95% CI +/- 31%) were better. After modification of the technique and optimization of image quality, both short term reproducibility and intraobserver variability improved (0% +/- 3%, 95% CI +/- 17% and 5% +/- 2%, 95% CI +/- 16% respectively). Aortic compliance can be measured using spin echo MR imaging with good reproducibility provided care is taken to obtain good quality images with high spatial resolution.

摘要

在47名健康志愿者中研究了磁共振成像(MR成像)测量主动脉顺应性的可重复性。测试了长期和短期可重复性以及观察者内变异性。对该方法进行了改进以提高图像质量,并再次测试了短期可重复性和观察者内变异性。为作比较,将自旋回波成像与电影梯度回波成像进行了对比。初始长期可重复性显示平均差异(±标准误)为3%(±7%),一致性界限的95%置信区间(CI)为±69%。短期可重复性(7%±6%,95%CI±46%)和观察者内变异性(1%±2%,95%CI±31%)较好。在对技术进行改进并优化图像质量后,短期可重复性和观察者内变异性均得到改善(分别为0%±3%,95%CI±17%和5%±2%,95%CI±16%)。只要注意获取具有高空间分辨率的高质量图像,使用自旋回波MR成像就可以很好地测量主动脉顺应性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验