Forbat S M, Mohiaddin R H, Yang G Z, Firmin D N, Underwood S R
Royal Brompton Hospital, London, UK.
J Magn Reson Imaging. 1995 Nov-Dec;5(6):635-9. doi: 10.1002/jmri.1880050604.
The reproducibility of MR imaging for the measurement of aortic compliance was studied in 47 healthy volunteers. Long and short term reproducibility and intraobserver variability were tested. The method was modified to improve image quality and short term reproducibility and intraobserver variability retested. For comparison, spin echo imaging was compared with cine gradient echo imaging. Initial long term reproducibility showed a mean difference (+/-SE) of 3% (+/- 7%) with 95% confidence interval (CI) for limits of agreement of +/- 69%. Short term reproducibility (7% +/- 6%, 95% CI +/- 46%) and intraobserver variability (1% +/- 2%, 95% CI +/- 31%) were better. After modification of the technique and optimization of image quality, both short term reproducibility and intraobserver variability improved (0% +/- 3%, 95% CI +/- 17% and 5% +/- 2%, 95% CI +/- 16% respectively). Aortic compliance can be measured using spin echo MR imaging with good reproducibility provided care is taken to obtain good quality images with high spatial resolution.
在47名健康志愿者中研究了磁共振成像(MR成像)测量主动脉顺应性的可重复性。测试了长期和短期可重复性以及观察者内变异性。对该方法进行了改进以提高图像质量,并再次测试了短期可重复性和观察者内变异性。为作比较,将自旋回波成像与电影梯度回波成像进行了对比。初始长期可重复性显示平均差异(±标准误)为3%(±7%),一致性界限的95%置信区间(CI)为±69%。短期可重复性(7%±6%,95%CI±46%)和观察者内变异性(1%±2%,95%CI±31%)较好。在对技术进行改进并优化图像质量后,短期可重复性和观察者内变异性均得到改善(分别为0%±3%,95%CI±17%和5%±2%,95%CI±16%)。只要注意获取具有高空间分辨率的高质量图像,使用自旋回波MR成像就可以很好地测量主动脉顺应性。