Cho Z H, Ro Y M, Park S T, Chung S C
Department of Radiological Sciences, University of California, Irvine 92717, USA.
Magn Reson Med. 1996 Jan;35(1):1-5. doi: 10.1002/mrm.1910350102.
The tailored radio frequency gradient echo (TRFGE) technique that has been used in venography (Cho et al., Magn. Reson. Med. 28, 25-38, 1992; Ro, Cho, Magn. Reson. Med. 28, 237, 1992) is applied to functional imaging. The TRFGE technique has the advantage that it is sensitive to the field gradient created by the susceptibility effect, thereby enhancing only the signals from the regions with local field gradient. In addition, the method is insensitive to the in-flow effect, especially from the arterial blood. The latter further simplifies the functional MRI data analysis. The TRFGE sequence is, therefore, suitable for functional MR imaging for which true susceptibility effect measurements are of prime importance. To examine the TRFGE functional imaging systematically, experiments with various imaging parameters such as flip angle alpha, repetition time, and echo time were performed and the results were compared with the data obtained from the conventional gradient echo functional MR imaging. The experimental results shown were all obtained from human volunteers with a 2.0T whole body MRI system.
已用于静脉造影的定制射频梯度回波(TRFGE)技术(Cho等人,《磁共振医学》,第28卷,第25 - 38页,1992年;Ro、Cho,《磁共振医学》,第28卷,第237页,1992年)被应用于功能成像。TRFGE技术的优势在于它对由磁化率效应产生的场梯度敏感,从而仅增强来自具有局部场梯度区域的信号。此外,该方法对流入效应不敏感,尤其是来自动脉血的流入效应。后者进一步简化了功能磁共振成像数据分析。因此,TRFGE序列适用于真正的磁化率效应测量至关重要的功能磁共振成像。为了系统地研究TRFGE功能成像,进行了各种成像参数(如翻转角α、重复时间和回波时间)的实验,并将结果与从传统梯度回波功能磁共振成像获得的数据进行比较。所示实验结果均来自使用2.0T全身磁共振成像系统的人类志愿者。