Vilser W, Schack B, Bareshova E, Senff I, Bräuer-Burchardt C, Münch K, Strobel J
Klinik für Augenkrankheiten der Universität Jena.
Ophthalmologe. 1995 Oct;92(5):728-34.
There are highly significant differences in the measuring results of arterial blood velocity between the indicator and laser-Doppler techniques (up to 800%). A new measuring procedure for the analysis of indicator dilution curves was developed based on indicator model and experimental results. The use of this new measuring procedure results in reduced mean systematic error between the indicator and laser-Doppler techniques to values around 10%. With the introduction of adaptive measuring arrays for the creation of indicator dilution curves and the application of adaptive algorithms for centering and spectral normalizing of the dilution curves, improved reproducibility can be expected.
指示剂法与激光多普勒技术在动脉血流速度测量结果上存在高度显著差异(高达800%)。基于指示剂模型和实验结果,开发了一种用于分析指示剂稀释曲线的新测量程序。使用这种新测量程序可将指示剂法与激光多普勒技术之间的平均系统误差降低至10%左右。随着用于创建指示剂稀释曲线的自适应测量阵列的引入以及用于稀释曲线定中心和频谱归一化的自适应算法的应用,有望提高测量的可重复性。