Newman V S, Berry J L, Routh W D, Ferrario C M, Dean R H
Bowman Gray School of Medicine, Wake Forest University, Winston-Salem, NC, USA.
J Vasc Interv Radiol. 1996 May-Jun;7(3):387-93. doi: 10.1016/s1051-0443(96)72877-4.
To compare the in vivo response to a new mechanically expandable vascular stent with the response to an existing type of balloon-expandable stent.
Prototype stents were deployed by means of a balloon catheter in the left iliac arteries of four healthy dogs. Palmaz stents were deployed in the contralateral iliac arteries to act as a control, and all stents were explanted after 6 weeks. Arteriography was performed at the time of insertion and before harvest, and pressure gradients were measured across each stent. The stents were then harvested and submitted for histologic examination.
The performance of the prototype stent was similar to that of the Palmaz stent with respect to structural integrity, migration, maintenance of intraluminal diameter, ease of deployment, radiopacity, and pressure gradients. Unlike the Palmaz stent, the prototype stent did not foreshorten during expansion. The stents showed a lack of uniformity in terms of the measured luminal area and neointimal accumulation. Neointimal accumulation was more confined to the struts of the prototype stent; the lumen therefore had a fluted appearance. Neointimal accumulation was more broadly distributed around the circumference of the vessel wall of the Palmaz stent.
In vivo performance of the prototype stent was similar to that of the Palmaz stent. Stent geometry may be an important determinant of neointimal response and resultant long-term patency.
比较新型机械可扩张血管支架与现有球囊可扩张支架的体内反应。
通过球囊导管将原型支架植入4只健康犬的左髂动脉。在对侧髂动脉植入帕尔马兹支架作为对照,6周后取出所有支架。在植入时和取出前进行血管造影,并测量每个支架两端的压力梯度。然后取出支架进行组织学检查。
在结构完整性、移位、管腔直径维持、植入难易程度、显影性和压力梯度方面,原型支架的性能与帕尔马兹支架相似。与帕尔马兹支架不同,原型支架在扩张过程中没有缩短。所测管腔面积和新生内膜堆积方面,支架表现出缺乏一致性。新生内膜堆积更多局限于原型支架的支柱;因此管腔呈有凹槽的外观。新生内膜堆积在帕尔马兹支架血管壁周围的圆周上分布更广泛。
原型支架的体内性能与帕尔马兹支架相似。支架几何形状可能是新生内膜反应和最终长期通畅性的重要决定因素。