使用[123I]碘安非他明和单光子发射计算机断层扫描对局部脑血流量定量进行的多中心验证。

A multicenter validation of regional cerebral blood flow quantitation using [123I]iodoamphetamine and single photon emission computed tomography.

作者信息

Iida H, Akutsu T, Endo K, Fukuda H, Inoue T, Ito H, Koga S, Komatani A, Kuwabara Y, Momose T, Nishizawa S, Odano I, Ohkubo M, Sasaki Y, Suzuki H, Tanada S, Toyama H, Yonekura Y, Yoshida T, Uemura K

机构信息

Department of Radiology and Nuclear Medicine, Research Institute for Brain and Blood Vessels, Akita, Japan.

出版信息

J Cereb Blood Flow Metab. 1996 Sep;16(5):781-93. doi: 10.1097/00004647-199609000-00003.

Abstract

Recently, two methods have been proposed for regional cerebral blood flow (rCBF) quantitation using [123I]iodoamphetamine (IMP) and single-photon emission computed tomography (SPECT). The table look-up (TLU) method has been shown to provide both rCBF and volume of distribution, Vd, images from two SPECT scans, while a single-scan autoradiographic (ARG) technique provided rCBF using a fixed and assumed Vd. In both methods, a single blood sample was referred to calibrate the previously determined standard input function. The present multicenter project was designed to evaluate the accuracy of both methods for use as clinical investigative tools. Ten independent institutions performed [123I]IMP-SPECT studies according to both methods in 76 subjects (10 normal volunteers, 32 patients with cerebrovascular disease, and 34 patients with other diseases). Calculated rCBF values were compared with those obtained by the following reference methods available in the participating institutions; [15O] H2O positron emission tomography (PET) (five institutions), [133Xe]SPECT (four institutions), and the [123I]IMP microsphere method (three institutions). Both ARG and TLU methods provided rCBF values that were significantly correlated with those measured by the [15O] H2O PET technique (p < 0.001 for all subjects; overall regression equation, y = 15.14 + 0.54x) and those measured by the [123I]IMP-microsphere method (p < 0.001 for all subjects: y = 2.0 + 0.80x). Significant correlation (p < 0.05) was observed in 18 of 24 subjects studied with the [133Xe] SPECT reference technique (overall regression equation, y = 15.0 + 0.55x). Mean cortical gray matter rCBF in a group of normal subject was 43.9 +/- 3.3 and 43.4 +/- 2.0 ml/min/100 g for the ARG and TLU methods, respectively. Regional Vd of [123I]IMP estimated by the TLU method was 45 ml/ml +/- 20% in the normal cortical region. Close agreement between ARG and TLU rCBF values was observed (y = -3.21 + 1.07x, r = 0.97), confirming the validity of assuming a fixed Vd in the ARG method. Results of this study demonstrate that both the ARG and TLU methods accurately and reliably estimate rCBF in a variety of clinical settings.

摘要

最近,有人提出了两种利用[123I]碘安非他明(IMP)和单光子发射计算机断层扫描(SPECT)定量局部脑血流量(rCBF)的方法。查表(TLU)法已被证明可从两次SPECT扫描中提供rCBF和分布容积Vd的图像,而单扫描放射自显影(ARG)技术则使用固定且假定的Vd来提供rCBF。在这两种方法中,均参考一份血样来校准先前确定的标准输入函数。本多中心项目旨在评估这两种方法作为临床研究工具的准确性。十个独立机构按照这两种方法对76名受试者(10名正常志愿者、32名脑血管疾病患者和34名其他疾病患者)进行了[123I]IMP-SPECT研究。将计算得到的rCBF值与参与机构中可用的以下参考方法所获得的值进行比较;[15O]H2O正电子发射断层扫描(PET)(五个机构)、[133Xe]SPECT(四个机构)以及[123I]IMP微球法(三个机构)。ARG法和TLU法提供的rCBF值与通过[15O]H2O PET技术测量的值(所有受试者p<0.001;总体回归方程,y = 15.14 + 0.54x)以及通过[123I]IMP微球法测量的值(所有受试者p<0.001:y = 2.0 + 0.80x)均显著相关。在用[133Xe]SPECT参考技术研究的24名受试者中的18名中观察到显著相关性(p<0.05)(总体回归方程,y = 15.0 + 0.55x)。一组正常受试者的平均皮质灰质rCBF,ARG法为43.9±3.3,TLU法为43.4±2.0 ml/min/100g。TLU法估计的正常皮质区域[123I]IMP的局部Vd为45 ml/ml±20%。观察到ARG法和TLU法的rCBF值高度一致(y = -3.21 + 1.07x,r = 0.97),证实了在ARG法中假定固定Vd的有效性。本研究结果表明,ARG法和TLU法在各种临床环境中均能准确可靠地估计rCBF。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验