Accuracy of the tonometric function of recycled gastric tonometer catheters.

作者信息

Freebairn R C, Buckley T A, Twyford C J, Oh T E

机构信息

Department of Anaesthesia and Intensive Care, Chinese University of Hong Kong, Prince of Wales Hospital, Shatin, N.T., Hong Kong.

出版信息

Anaesth Intensive Care. 1995 Oct;23(5):570-3. doi: 10.1177/0310057X9502300507.

Abstract

We investigated the accuracy of recycled gastric tonometer catheters, using a prospective laboratory model, within the Intensive Care Unit of a University teaching hospital. Ten used tonometer catheters and three new catheters were exposed to known constant PCO2 levels between 23-65 torr (3.06-8.66 kPa) in a gas mixing chamber, at a temperature of 35.5-38 degrees C and a relative humidity > 95%. The tonometer balloons were primed with normal saline, and after a 30-minute equilibration time, the saline was sampled for PCO2 measurements. Steady-state PCO2 (PCO2(SS)) and gastric mucosal pH (pHi) values were derived using a correction factor and the Henderson-Hasselbalch equation. There was no difference in the mean or distribution of PCO2(SS) values between recycled and new catheters. The PCO2(SS) values of recycled catheters were within the 95% confidence limits of those of new catheters. Individual catheter values did not differ significantly from each other (ANOVA P = 0.05). Using an arterial HCO3- value of 21 mmol/l, the calculated pHi values of used catheters varied < 1.5% of the mean pHi of new catheters. The tonometric performance of recycled gastric tonometer catheters was similar to that of new catheters when tested in vitro.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验