Hurkmans C, Knöös T, Nilsson P
Department of Radiation Physics, University Hospital in Lund, Sweden.
Phys Med Biol. 1996 Aug;41(8):1277-90. doi: 10.1088/0031-9155/41/8/003.
Output normalized dose profiles for asymmetric open photon fields has been calculated using a commercial treatment planning system (TPS) based on a dose-to-energy-fluence concept. The model does not require any additional measurements for off-axis fields. Calculations are compared with measurements for quadratic fields of 5 cm x 5 cm up to 20 cm x 20 cm, with their geometric field centre positioned 10 cm off-axis in the in-plane direction. The measurements include depth doses and profiles in-plane as well as cross-plane for nominal photon energies of 4, 6 and 18 MV x-rays. Both calculated and measured doses are normalized with respect to a 10 cm x 10 cm reference field, therefore making it possible to compare not only the relative distributions but also the absolute dose levels; that is, calculation of monitor units is included. The calculated depth-dose curves are generally in good agreement with measured data with an accuracy at the absolute dose level of 2% at depths beyond the dose maximum. The cross-plane profiles are calculated with an accuracy better than 3% within the field. The 'tilt' towards the collimator central axis of the in-plane profiles is predicted by the model, but is somewhat overestimated at large depths. The system provides the possibility to separate the primary and scattered parts of the dose and the cause of this tilting was studied by comparing calculated phantom-scattering and head-scattering dose profiles for a symmetric 40 cm x 20 cm field to dose profiles for an asymmetric 20 cm x 20 cm field. The tilting is shown to originate from a change both in phantom scattering and in head scattering compared to the case of symmetrical fields. The results indicate that the investigated TPS can calculate dose distributions in open asymmetric fields with a high degree of accuracy, typically better than 2-3%.
已使用基于剂量与能量注量概念的商用治疗计划系统(TPS)计算了非对称开放光子野的归一化剂量分布。该模型不需要对离轴野进行任何额外测量。将计算结果与5 cm×5 cm至20 cm×20 cm二次野的测量结果进行比较,其几何野中心在平面内方向离轴10 cm。测量包括4、6和18 MV X射线标称光子能量下的深度剂量、平面内以及平面交叉方向的剂量分布。计算剂量和测量剂量均相对于10 cm×10 cm参考野进行归一化,因此不仅可以比较相对分布,还可以比较绝对剂量水平;即包括监测单位的计算。计算得到的深度剂量曲线与测量数据总体吻合良好,在剂量最大值以外的深度处,绝对剂量水平的精度为2%。平面交叉方向的剂量分布计算精度在野内优于3%。模型预测了平面内剂量分布向准直器中心轴的“倾斜”,但在大深度处有些高估。该系统提供了分离剂量的原发射线和散射部分的可能性,并通过比较对称40 cm×20 cm野的计算体模散射和机头散射剂量分布与非对称20 cm×20 cm野的剂量分布,研究了这种倾斜的原因。结果表明,与对称野情况相比,这种倾斜源于体模散射和机头散射的变化。结果表明,所研究的TPS能够高精度地计算开放非对称野中的剂量分布,通常优于2% - 3%。