Intraluminal micromanometry: an evaluation of the dynamic performance of micro-extrusions and sleeve sensors.

作者信息

Omari T, Bakewell M, Fraser R, Malbert C, Davidson G, Dent J

机构信息

Gastroenterology Unit, Women's and Children's Hospital, Adelaide, Australia.

出版信息

Neurogastroenterol Motil. 1996 Sep;8(3):241-5. doi: 10.1111/j.1365-2982.1996.tb00263.x.

Abstract

Conventional manometric techniques are unsuitable for studies in premature infants and small laboratory animals. We have therefore developed silicone rubber 5-lumen and 10-lumen micromanometric extrusions with an o.d. 2.0 mm and lumina of 0.35 mm i.d. This study evaluates the suitability of microextrusions for intraluminal perfusion manometry. Pressure offset, post-occlusion pressure rise rate and sphincter model studies were used to assess the manometric performance of the extrusions and a miniature sleeve sensor (25 mm long) at infusion rates of 0.01-0.1 mL min-1. Micro-extrusions (5-lumen/10-lumen, respectively) had offsets (per 100 cm of length) of 3.8/5.0 mmHg at 0.01 mL min-1 and 25.6/26.2 mmHg at 0.1 mL min-1 and rise rates (in 160 cm lengths) of 64/43 mmHg sec-1 at 0.01 mL min-1 and 330/224 mmHg sec-1 at 0.1 mL min-1. Infusion rates 0.025 mL min-1 produced rise rates 100 mmHg sec-1. The miniature sleeve sensor had minimal resistance to perfusion, rise rates of 3 mmHg sec-1 at 0.01 mL min-1 and 23 mmHg sec-1 at 0.1 mL min-1 and recorded pressure as accurately as a side hole. We conclude that the performance of micromanometric extrusions and sleeves is sufficient for intraluminal perfusion manometry.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验