Yew O F, Cheong Y M, Zaidatul A R, Thanaletchimy N
Division of Bacteriology, Institute for Medical Research, Kuala Lumpur.
Malays J Pathol. 1995 Jun;17(1):11-5.
The BACTEC NR730 blood culture system is a commercially available system which utilizes infrared spectrophotometry to detect bacterial growth in broth media inoculated with the blood specimen. The objective of this study is to compare this system with the conventional system used in our laboratory. A total of 479 blood samples were collected from adult patients in the Klang TAR Hospital and inoculated simultaneously into the two systems. Clinically significant isolates were recovered from 59 (12.3%) specimens by one or both systems yielding a total of 65 isolates of which 41 isolates were recovered in both systems, 18 in BACTEC only and 6 in the conventional system (P < 0.01). Of the 40 (8.3%) blood cultures that were judged as contaminated, 2 were found in both systems, 12 in the BACTEC only and 26 in the conventional system. After 48 hours (day 2) the BACTEC system had recovered 93.2% of the total significant isolates while the conventional system recovered 61.7% only. Our results showed that the BACTEC system yielded more significant isolates in a shorter time and with a lower incidence of contamination as compared to the conventional system.
BACTEC NR730血培养系统是一种市售系统,它利用红外分光光度法检测接种了血标本的肉汤培养基中的细菌生长情况。本研究的目的是将该系统与我们实验室使用的传统系统进行比较。从巴生塔医院的成年患者中总共采集了479份血样,并同时接种到这两种系统中。通过一种或两种系统从59份(12.3%)标本中分离出具有临床意义的菌株,共获得65株分离株,其中41株在两种系统中均被分离出,18株仅在BACTEC系统中被分离出,6株在传统系统中被分离出(P<0.01)。在被判定为污染的40份(8.3%)血培养物中,两种系统中均发现2份,仅在BACTEC系统中发现12份,在传统系统中发现26份。48小时(第2天)后,BACTEC系统分离出了93.2%的全部有意义分离株,而传统系统仅分离出61.7%。我们的结果表明,与传统系统相比,BACTEC系统能在更短的时间内分离出更多有意义的菌株,且污染发生率更低。