Suppr超能文献

一种用于测定血清中甲状腺素的自动竞争性结合方法。

An automated competitive binding procedure for measuring thyroxine in serum.

作者信息

Hofman L F, Bouley A M, Barron E J

出版信息

Clin Chem. 1977 Sep;23(9):1628-32.

PMID:890905
Abstract

A competitive binding assay for serum thyroxine has been automated, with the use of small, reusable Sephadex columns to separate thyroxine from endogenous thyroid-binding globulin and later to separate the bound and free thyroxine. Sixty samples an hour are run through columns, which are held in an aluminum turntable rotated by a fraction collector motor. Reagents and samples are fed to the columns by a proportioning pump. Waste eluates are collected and drained to the sink by a Teflon tray positioned between the columns and counting tubes, also held by the turntable. A cut out area in the tray allows one to collect the bound fraction, which can then be counted in a scintillation counter. Values are calculated and printed by a desk calculator interfaced with the scintillation counter. The "day-to-day" CV for this method is 5%. The accuracy is satisfactory when tested by comparison with other methods, by recoveries, and by linearity of dilutions.

摘要

一种用于血清甲状腺素的竞争性结合测定法已实现自动化,该方法使用小型可重复使用的葡聚糖凝胶柱,先将甲状腺素与内源性甲状腺结合球蛋白分离,随后再分离结合型和游离型甲状腺素。每小时有60个样本通过柱子,这些柱子放置在由分部收集器电机驱动旋转的铝制转盘上。试剂和样本由比例泵输送到柱子中。废液通过位于柱子和计数管之间的聚四氟乙烯托盘收集并排到水槽中,该托盘也由转盘固定。托盘中的一个切口区域可用于收集结合部分,然后可在闪烁计数器中进行计数。数值由与闪烁计数器连接的台式计算器计算并打印出来。该方法的“日常”变异系数为5%。通过与其他方法比较、回收率测试以及稀释线性测试,其准确性令人满意。