Transmission pulse oximetry in the fetal lamb: is there a universal calibration?

作者信息

Nijland R, Jongsma H W, Crevels J, Menssen J J, Nijhuis J G, Oeseburg B

机构信息

Department of Obstetrics and Gynaecology, University of Nijmegen, The Netherlands.

出版信息

Pediatr Res. 1996 Mar;39(3):464-9. doi: 10.1203/00006450-199603000-00014.

Abstract

Transmission pulse oximetry is widely used for oxygen monitoring. The use of pulse oximeters is steadily expanding toward situations with low arterial oxygen saturation (Sao2) values. Therefore, we evaluated transmission pulse oximetry in the unanesthetized fetal lamb at low Sao2 levels. In seven fetal lambs, fetal hypoxemia was induced by occlusion of the maternal common iliac artery, four days after the instrumentation of the animal. Two Nellcor prototype transmission Y-sensors (light emitting diodes: 660 and 890 nm) were applied, one around a forelimb muscle and one around a skinfold in the neck, and were connected to Nellcor pulse oximeters. The pulse oximeter was calibrated for the skin measurements. Pulse oximeter saturation readings (Spo2) were compared with sample Sao2 values, over an Sao2 range of 13 to 63%. For the neck sensor the SD of the difference was 5.0% (n = 101). For the muscle sensor the mean difference was 19.5% and the SD of the difference was 5.9% (n = 206). Regression analysis showed a different calibration line for the muscle sensor with the equation: Spo2 = 0.92 x Sao2 + 21.90. Continuous recordings were obtained both from the forelimb muscle and from the neck, but the recordings from the neck showed a substantial loss of signal during the hypoxemia period. We conclude that transmission pulse oximetry is less accurate below an Sao2 of 70% in fetal lambs than above 70% Sao2. At these low levels of Sao2, pulse oximeters may need to be constructed with different calibration lines for various application positions of the body.

摘要

透射式脉搏血氧饱和度测定法广泛用于氧监测。脉搏血氧仪的应用正稳步扩展至动脉血氧饱和度(Sao2)值较低的情况。因此,我们评估了未麻醉的胎羊在低Sao2水平下的透射式脉搏血氧饱和度测定法。在7只胎羊中,在动物植入仪器4天后,通过阻断母羊的双侧髂总动脉诱导胎儿低氧血症。应用了两个Nellcor原型透射式Y传感器(发光二极管:660和890纳米),一个环绕前肢肌肉,一个环绕颈部的皮肤褶,然后连接到Nellcor脉搏血氧仪。对用于皮肤测量的脉搏血氧仪进行了校准。在13%至63%的Sao2范围内,将脉搏血氧仪的饱和度读数(Spo2)与样本Sao2值进行比较。对于颈部传感器,差值的标准差为5.0%(n = 101)。对于肌肉传感器,平均差值为19.5%,差值的标准差为5.9%(n = 206)。回归分析显示肌肉传感器的校准线不同,方程为:Spo2 = 0.92×Sao2 + 21.90。从前肢肌肉和颈部都获得了连续记录,但颈部记录在低氧血症期间出现了大量信号丢失。我们得出结论,在胎羊中,透射式脉搏血氧饱和度测定法在Sao2低于70%时比在Sao2高于70%时准确性更低。在这些低Sao2水平下,脉搏血氧仪可能需要针对身体的不同应用部位构建不同的校准线。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验