Dassel A C, Graaff R, Aarnoudse J G, Elstrodt J M, Heida P, Koelink M H, de Mul F F, Greve J
Department of Obstetrics and Gynecology, University Hospital Groningen, The Netherlands.
Pediatr Res. 1992 Mar;31(3):266-9. doi: 10.1203/00006450-199203000-00014.
Transmission pulse oximetry is used for monitoring in many clinical settings. However, for fetal monitoring during labor and in situations with poor peripheral perfusion, transmission pulse oximetry cannot be used. Therefore, we developed a reflectance pulse oximeter, which uses the relative intensity changes of the reflected red and infrared light (red/infrared ratio) to measure the arterial oxygen saturation. The performance of the reflectance pulse oximeter was studied in acute experiments in fetal lambs. By stepwise reduction of the inspired oxygen concentration of the ewe, measurements were done at the fetal scalp at various arterial oxygen saturation levels (17-82%). Reflectance pulse oximeter readings were averaged over periods of 15 s and compared with simultaneously taken fetal arterial blood samples. A calibration curve for the relationship between red/infrared ratio and arterial oxygen saturation was obtained from 53 measurements in four fetal lambs, by linear regression analysis [red/infrared = 4.088-(0.038.SaO2), r = 0.96]. In these experiments, the pulse oximeter showed a precision of 4.7% oxygen saturation around the calibration curve, with a 95% confidence interval of +/- 9.4%.
透射式脉搏血氧饱和度仪在许多临床环境中用于监测。然而,在分娩期间的胎儿监测以及外周灌注不良的情况下,不能使用透射式脉搏血氧饱和度仪。因此,我们开发了一种反射式脉搏血氧仪,它利用反射红光和红外光的相对强度变化(红/红外比)来测量动脉血氧饱和度。在胎儿羔羊的急性实验中研究了反射式脉搏血氧仪的性能。通过逐步降低母羊的吸入氧浓度,在不同动脉血氧饱和度水平(17 - 82%)下于胎儿头皮处进行测量。反射式脉搏血氧仪读数在15秒的时间段内进行平均,并与同时采集的胎儿动脉血样本进行比较。通过线性回归分析,从四只胎儿羔羊的53次测量中获得了红/红外比与动脉血氧饱和度之间关系的校准曲线[红/红外 = 4.088 - (0.038·SaO₂),r = 0.96]。在这些实验中,脉搏血氧仪在校准曲线周围显示出4.7%血氧饱和度的精度,95%置信区间为±9.4%。