Initial clinical experience with the Peacock intensity modulation of a 3-D conformal radiation therapy system.

作者信息

Carol M, Grant W H, Pavord D, Eddy P, Targovnik H S, Butler B, Woo S, Figura J, Onufrey V, Grossman R, Selkar R

机构信息

Methodist Hospital, Baylor College of Medicine, Houston, Tex, USA.

出版信息

Stereotact Funct Neurosurg. 1996;66(1-3):30-4. doi: 10.1159/000099664.

Abstract

Peacock is a 3-D conformal treatment planning and delivery system for conformal radiation therapy which delivers intensity-modulated fields. A group of 13 patients were treated between March and February, 1995. Patient age ranged from 10 to 74. Six of the patients had previously received radiation therapy. Target volume was from 2.5 to 70 cm3: all treatments were fractionated, using single table angle plan with 270 degrees of rotation. A removable invasive cranial screw fixation device was used in all cases. Treated isodose line averaged 85%; conformality index was better than 2, with 0-5% of critical structure volume exceeding dose limits. Setup time averaged 8 min per fraction: treatment time ranged from 10 to 80 min. There were two equipment failures in 200 fractions/1,000 gantry rotations. Patient position averaged about 1 mm within initial setup. Follow-up has revealed no complications from the radiation.

摘要

孔雀系统是一种用于适形放射治疗的三维适形治疗计划与实施系统,可提供调强射野。1995年3月至次年2月期间,对一组13例患者进行了治疗。患者年龄在10岁至74岁之间。其中6例患者此前接受过放射治疗。靶体积为2.5至70立方厘米:所有治疗均采用分割照射,使用单野角度计划,机架旋转270度。所有病例均使用了可移除的侵入性颅骨螺钉固定装置。治疗等剂量线平均为85%;适形指数大于2,关键结构体积超过剂量限值的比例为0 - 5%。每次分割的摆位时间平均为8分钟:治疗时间为10至80分钟。在200次分割/1000次机架旋转中有两次设备故障。患者在初始摆位后的位置平均偏差约1毫米。随访结果显示未出现放射相关并发症。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验