皮肤血流的激光多普勒测量。

Laser Doppler measurement of cutaneous blood flow.

作者信息

Holloway G A, Watkins D W

出版信息

J Invest Dermatol. 1977 Sep;69(3):306-9. doi: 10.1111/1523-1747.ep12507665.

Abstract

This work describes an instrument for the noninvasive measurement of cutaneous blood flow velocity. The system utilizes the Doppler shift of laser light backscattered from moving red blood cells in the cutaneous microcirculation, the shift being obtained by an optical heterodyning technique. Comparison is made between this technique and the 133xenon clearance technique in measuring cutaneous flow in the forearms of normal volunteers. Variations in flow were obtained by inducing different degrees of solar erythema with an ultraviolet sunlamp. A Y on X linear regression yielded a regression coefficient = 0.89 (p less than 0.001, n = 16) between the two methods. The laser Doppler method appear to represent a practical technique for clinical evaluation of cutaneous blood flow in any skin surface.

摘要

这项工作描述了一种用于无创测量皮肤血流速度的仪器。该系统利用激光在皮肤微循环中移动的红细胞后向散射产生的多普勒频移,这种频移通过光学外差技术获得。在正常志愿者的前臂测量皮肤血流时,对该技术与133氙清除技术进行了比较。通过用紫外线太阳灯诱导不同程度的日晒红斑来获得血流变化。两种方法之间的X对Y线性回归得出回归系数=0.89(p<0.001,n=16)。激光多普勒方法似乎是一种用于临床评估任何皮肤表面皮肤血流的实用技术。