Experimental evaluation of a newly developed ultrathin silicone layer coated hollow fiber oxygenator.

作者信息

Shimono T, Shomura Y, Tahara K, Hioki I, Tenpaku H, Maze Y, Hirano R, Shimpo H, Shionoya Y, Yokoyama A, Morikan T, Yada I

机构信息

Department of Thoracic and Cardiovascular Surgery, Mie University, School of Medicine, Japan.

出版信息

ASAIO J. 1996 Sep-Oct;42(5):M451-4. doi: 10.1097/00002480-199609000-00029.

Abstract

The authors developed a new membrane oxygenator that consists of microporous polypropylene hollow fibers coated with a 0.2 micron ultrathin silicone layer. Five venoarterial bypasses were conducted on mongrel dogs for 24 hr using these new oxygenators. The blood flow rate was maintained at 750 ml/min, and the V/Q ratio was maintained at 1:1. As a control, three venoarterial bypasses were conducted under the same conditions using an oxygenator with the same design but without the silicone coating. Eight to 16 hr after the initiation of bypass, severe plasma leakage occurred in all control experiments, so the bypasses were terminated. However, plasma leakage did not occur throughout the 24 hr of any of the experiments using the new oxygenator. The O2 transfer rate of the new oxygenators after 24 hr of perfusion was 59.7 +/- 6.6 ml/min/m2, and the plasma free hemoglobin level 8 hr after the initiation of bypass was 41.4 +/- 40.2 mg/dl, compared with 145.3 +/- 189.6 mg/dl in the control group. Scanning electron microscopic examination of the silicone coated fibers after 24 hr of bypass revealed a few scattered platelet adherents and no damage to the silicone coated surface. These results suggest that this new oxygenator has satisfactory gas transfer and good durability.

摘要

作者研发了一种新型膜式氧合器,它由涂有0.2微米超薄硅胶层的微孔聚丙烯中空纤维组成。使用这些新型氧合器对杂种犬进行了5次静脉-动脉体外循环,持续24小时。血流速度维持在750毫升/分钟,通气/血流比值维持在1:1。作为对照,在相同条件下使用设计相同但无硅胶涂层的氧合器进行了3次静脉-动脉体外循环。体外循环开始后8至16小时,所有对照实验均出现严重的血浆渗漏,因此终止了体外循环。然而,使用新型氧合器的任何实验在整个24小时内均未发生血浆渗漏。灌注24小时后,新型氧合器的氧转运率为59.7±6.6毫升/分钟/平方米,体外循环开始8小时后的血浆游离血红蛋白水平为41.4±40.2毫克/分升,而对照组为145.3±189.6毫克/分升。体外循环24小时后对硅胶涂层纤维进行扫描电子显微镜检查,发现有一些散在的血小板黏附物,硅胶涂层表面未受损。这些结果表明,这种新型氧合器具有令人满意的气体传输性能和良好的耐用性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验