Abstract

Epididymal sperm counts, a common measurement in male reproductive toxicity studies, are routinely determined using a hemacytometer. Recently, computer assisted methods for automated sperm counts have been developed. In the present study we evaluated an automated system, the TOX IVOS (Hamilton Thorne Research, Beverly, MA) HTM-IDENT option, that utilizes a DNA-specific stain and fluorescence illumination to identify sperm for enumeration. Cauda and caput epididymal sperm counts were determined in 48 adult male Sprague-Dawley rats, using both the hemacytometer and HTM-IDENT. The mean hemacytometer and HTM-IDENT counts (+/- SD) were 250 +/- 43 and 254 +/- 52 million, respectively, for cauda sperm, and 123 +/- 13 and 127 +/- 18 million, respectively, for caput sperm. The average coefficient of variation using the hemacytometer was 13.8% as compared to 17.3% for the HTM-IDENT. Comparison of the machine count and a visual count from the Display Statics screen of the HTM-IDENT indicated that when two or more sperm heads touched or overlapped, the machine counted them as one. Manual (visual) and machine counts when compared over a range of nine concentrations from 3.7 to 47.8 million/mL differed by 4 to 12% at the lowest to highest concentration. The concentration of epididymal sperm samples used in comparing the two counting methods ranged from 5.8 to 17.7 million/mL. Therefore, the HTM-IDENT undercounting error attributable to sperm heads touching was less than 6%. Overall the data indicate good agreement between the HTM-IDENT and the hemacytometer counts. Furthermore, both counting time and technician fatigue were markedly reduced. Thus the HTM-IDENT option improves the efficiency of epididymal sperm counting without loss of precision.

摘要

附睾精子计数是雄性生殖毒性研究中的一项常见测量指标,通常使用血细胞计数器来测定。最近,已开发出用于自动精子计数的计算机辅助方法。在本研究中,我们评估了一种自动化系统——TOX IVOS(汉密尔顿·桑恩研究公司,马萨诸塞州贝弗利)的HTM-IDENT选项,该系统利用一种DNA特异性染色剂和荧光照明来识别精子以便计数。使用血细胞计数器和HTM-IDENT对48只成年雄性斯普拉格-道利大鼠的附睾尾部和头部精子进行计数。附睾尾部精子的血细胞计数器平均计数(±标准差)为2.50±0.43亿,HTM-IDENT平均计数为2.54±0.52亿;附睾头部精子的血细胞计数器平均计数为1.23±0.13亿,HTM-IDENT平均计数为1.27±0.18亿。使用血细胞计数器的平均变异系数为13.8%,而HTM-IDENT为17.3%。对HTM-IDENT显示统计屏幕上的机器计数和目视计数进行比较表明,当两个或更多精子头部接触或重叠时,机器将它们计为一个。在370万至4780万/mL的九个浓度范围内进行比较时,手动(目视)计数和机器计数在最低浓度至最高浓度下相差4%至12%。用于比较两种计数方法的附睾精子样本浓度范围为580万至1770万/mL。因此,由于精子头部接触导致的HTM-IDENT计数不足误差小于6%。总体而言,数据表明HTM-IDENT和血细胞计数器计数之间具有良好的一致性。此外,计数时间和技术人员的疲劳都明显减少。因此,HTM-IDENT选项提高了附睾精子计数的效率,且不损失精度。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验