Wang S S, Chu S H, Chou N K, Qian K X
Department of Surgery, National Taiwan University, Taipei.
Artif Organs. 1996 Dec;20(12):1310-3. doi: 10.1111/j.1525-1594.1996.tb00681.x.
Because of severe hemolysis, especially on producing pulsatile flow by changing the rotating speed of the impellers, the traditional centrifugal pump was rarely used for long-term support of the failing heart. We therefore developed a motor driven pulsatile implantable impeller pump. The pulsatility was achieved by changing the rotating speed via introducing a square waveform voltage into the motor coil. The impeller vane was designed to have both radial and axial curves according to the stream surface and stream lines to reduce the thrombosis and hemolysis. Nine calves weighing 80 to 100 kg were used. With the calves under endotracheal general anesthesia, left posterolateral thoracotomy was performed to connect the inflow tube with the left atrial appendage and to anastomose the outflow tube with the descending aorta. The calves usually awoke and stood up within hours after discontinuation of anesthetics. Within 7 days, continuous monitoring of electrocardiogram, systemic and pulmonary arterial pressures, and central venous pressure were performed to adjust the pump flow to 40% to 50% of the cardiac output. During the survival of 4 to 54 days (mean 16.3 +/- 19.3 days with two calves surviving longer than 1 month), no significant deterioration of liver or renal function was noted. Because of bleeding, hemoglobin reduced from 11.4 +/- 1.8 to 9.0 +/- 1.3 g/dl, and the hematocrit decreased from 34.5 +/- 4.7 to 26.7 +/- 4.6%. No significant changes of free hemoglobin were noted. In our results, the device revealed competent pulsatile function without severe blood damage or organ dysfunction.
由于严重溶血,尤其是通过改变叶轮转速来产生脉动血流时,传统离心泵很少用于长期支持衰竭心脏。因此,我们研发了一种电动驱动的可植入式脉动叶轮泵。通过向电机线圈引入方波电压来改变转速,从而实现脉动性。根据流面和流线,将叶轮叶片设计为具有径向和轴向曲线,以减少血栓形成和溶血。使用了9头体重80至100公斤的小牛。在气管内全身麻醉下对小牛进行左后外侧开胸手术,将流入管与左心耳相连,并将流出管与降主动脉吻合。小牛通常在停止麻醉后数小时内苏醒并站立起来。在7天内,持续监测心电图、体循环和肺动脉压以及中心静脉压,将泵流量调整至心输出量的40%至50%。在4至54天的存活期内(平均16.3±19.3天,两头小牛存活时间超过1个月),未观察到肝功能或肾功能有明显恶化。由于出血,血红蛋白从11.4±1.8降至9.0±1.3克/分升,血细胞比容从34.5±4.7降至26.7±4.6%。未观察到游离血红蛋白有明显变化。在我们的结果中,该装置显示出良好的脉动功能,没有严重的血液损伤或器官功能障碍。