Water and bovine serum lubrication compared in simulator PTFE/CoCr wear model.

作者信息

Good V D, Clarke I C, Anissian L

机构信息

Howard and Irene Peterson Tribology Laboratory, Department of Orthopaedics, Loma Linda University Medical Center, California 92354, USA.

出版信息

J Biomed Mater Res. 1996 Winter;33(4):275-83. doi: 10.1002/(SICI)1097-4636(199624)33:4<275::AID-JBM7>3.0.CO;2-M.

Abstract

Controversy surrounds wear data from laboratory hip simulator studies, whether derived from water-based or serum-based studies or whether a major design parameter such as the size of the femoral head has an effect on the volume of wear particulate released. To investigate these relationships, we studied cup wear in water- and serum-based lubricants using as our standard the polytetrafluoroethylene (PTFE) data derived by Charnley. To model Charnley's clinical experience, PTFE acetabular cups were used in sets of three each of four sizes of CoCr femoral heads: 22.25-, 28-, 32-, and 42-mm diameters. Six criteria were used to evaluate the performance of the lubricants against clinical accuracy and scientific methods. The PTFE wear data from the serum-based tests was consistently linear with the duration of the test, exhibited a precision within +/-3% about the average for each set of three cups, and copious amounts of wear debris were clearly seen circulating and settling to the bottom of the wear chambers. The wear data clearly demonstrated Charnley's thesis that volume of wear increases with regard to the size of the femoral head in a linear manner. This increase was considered satisfactory at 9%/ mm. However, in terms of clinical accuracy, the simulator wear rates averaged 3 to 4 times greater than the comparable clinical data for wear magnitude. Thus, the serum-based tests satisfied three of the six criteria used. The PTFE wear data from the water-based tests was generally nonlinear, continually increasing with test duration. These wear trends were examined in three discrete phases to estimate the changing wear rates. By the end of the tests, the wear rates had increased from 1.3 to 3.9 times, with the 42-mm heads showing the greatest change. The resulting precision was never better than +/-26% and deteriorated to +/-70%. In terms of clinical accuracy, the water-based wear rates varied from 2 to 7 times less than the Charnley PTFE wear magnitudes, averaging 4 times less. The water-based data did not satisfactorily model the relationship between increased wear with increased head size. Minimal PTFE wear debris was observed, and what did emerge after thousands of wear cycles appeared as streamers up to 30 mm long and up to 5 mm wide. When these detached, they floated up to the surface where they could be separated into smaller particulates. A similar phenomenon was noted for polyethylene wear tests conducted with water lubrication. Thus the water-based tests satisfied none of the six validation criteria evaluated. These data raise serious doubts as to the validity of testing implant and material combinations in water as a predictor of clinical performance. Bovine serum was not totally satisfactory, but the wear data did model some of the important clinical characteristics of hip joint behavior.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验