[Evaluation of vaporizers by anesthetic gas monitors corrected with a new method for preparation of calibration gases].

作者信息

Kurashiki T

机构信息

Department of Anesthesiology, San-in Rosai Hospital, Yonago.

出版信息

Masui. 1996 Nov;45(11):1354-66.

PMID:8953869
Abstract

For resolving the discrepancy of concentrations found among anesthetic gas monitors, the author proposed a new method using a vaporizer as a standard anesthetic gas generator for calibration. In this method, the carrier gas volume is measured by a mass flow meter (SEF-510 + FI-101) installed before the inlet of the vaporizer. The vaporized weight of volatile anesthetic agent is simultaneously measured by an electronic force balance (E12000S), on which the vaporizer is placed directly. The molar percent of the anesthetic is calculated using these data and is transformed into the volume percent. These gases discharging from the vaporizer are utilized for calibrating anesthetic gas monitors. These monitors are normalized by the linear equation describing the relationship between concentrations of calibration gases and readings of the anesthetic gas monitors. By using normalized monitors, flow rate-concentration performance curves of several anesthetic vaporizers were obtained. The author concludes that this method can serve as a standard in evaluating anesthetic vaporizers.

摘要

相似文献

1
[Evaluation of vaporizers by anesthetic gas monitors corrected with a new method for preparation of calibration gases].
Masui. 1996 Nov;45(11):1354-66.
2
Refractive indices for volatile anesthetic gases: equipment and method for calibrating vaporizers and monitors.挥发性麻醉气体的折射率:校准蒸发器和监测仪的设备及方法
J Clin Monit. 1995 May;11(3):168-74. doi: 10.1007/BF01617718.
3
Comparison of use of an infrared anesthetic gas monitor and refractometry for measurement of anesthetic agent concentrations.
Am J Vet Res. 2011 Oct;72(10):1299-304. doi: 10.2460/ajvr.72.10.1299.
4
Technical communication: inhaled anesthetic agent-vaporizer mismatch: management in settings with limited resources: don't try this at home.技术交流:吸入麻醉药-蒸发器不匹配:资源有限环境下的管理:切勿在家尝试。
Anesth Analg. 2013 Jun;116(6):1272-5. doi: 10.1213/ANE.0b013e31828d6aeb. Epub 2013 Apr 4.
5
Pollution of ambient air by volatile anesthetics: a comparison of 4 anesthetic management techniques.挥发性麻醉剂对环境空气的污染:四种麻醉管理技术的比较
AANA J. 2006 Apr;74(2):121-5.
6
[A proposal for the control of anesthetic gas vaporizers].
Ugeskr Laeger. 1991 Nov 25;153(48):3415-6.
7
Cardiorespiratory effects of low-flow and closed circuit inhalation anesthesia, using sevoflurane delivered with an in-circuit vaporizer and concentrations of compound A.使用回路内蒸发器输送七氟烷及不同浓度的化合物A进行低流量和闭路吸入麻醉时的心肺效应
Am J Vet Res. 1998 May;59(5):603-8.
8
Evaluation of isoflurane and sevoflurane vaporizers over a wide range of oxygen flow rates.在广泛的氧气流速范围内对异氟烷和七氟烷蒸发器进行评估。
Am J Vet Res. 2006 Jun;67(6):936-40. doi: 10.2460/ajvr.67.6.936.
9
Advanced technologies and devices for inhalational anesthetic drug dosing.用于吸入麻醉药给药的先进技术和设备。
Handb Exp Pharmacol. 2008(182):451-70. doi: 10.1007/978-3-540-74806-9_21.
10
The importance of monitoring volatile anesthetic concentration.监测挥发性麻醉剂浓度的重要性。
J Clin Anesth. 2018 Nov;50:2-4. doi: 10.1016/j.jclinane.2018.06.031. Epub 2018 Jun 19.