Thomsen T A, Hokanson E, Barnes R W
Med Instrum. 1977 Jul-Aug;11(4):240-3.
Strain gauge plethysmography has permitted noninvasive quantitation of altered venous hemodynamics in the legs of patients with venous disease. A new plethysmograph has been developed which avoids previous problems of mechanical calibration of gauges and the necessity for a recorder for calculating blood flows. This instrument permits accurate electrical calibration of mercury-in-Silastic gauges which are effectively at the corners of the Wheatstone bridge and are supplied by constant current for linear output. A timer and follow-and-hold circuit allow automatic calculation of rates of change in limb circumference and, thus, blood flow, from a panel meter. The accuracy of the automatic versus the graphic methods of quantitating venous outflow and reflux flow was determined in 20 normal subjects. Venous outflows were 15.1 +/- 0.5 (mean +/- 1 S.E.M.) and 15.9 +/- 0.4 cc/min/100 cc by automatic and graphic methods, respectively (r = 0.96, p less than 0.01). Reflux flows were 2.1 +/- 0.2 and 2.3 +/- 0.2 cc/min/100 cc by the automatic and graphic methods, respectively (r = 0.93, p less than 0.01). This plethysmograph may prove to be a simple, rapid, and accurate device for evaluating altered hemodynamics in venous disease.
应变片体积描记法已能够对静脉疾病患者腿部静脉血流动力学的改变进行无创定量分析。现已研制出一种新型体积描记仪,它避免了以往应变片机械校准的问题以及计算血流时需要记录仪的必要性。该仪器能对硅橡胶汞应变片进行精确的电校准,这些应变片实际上位于惠斯通电桥的角上,并由恒流供电以获得线性输出。一个定时器和跟踪保持电路可根据面板仪表自动计算肢体周长的变化率,从而计算出血流。在20名正常受试者中测定了自动法与图形法在定量静脉流出量和反流流量方面的准确性。自动法和图形法测得的静脉流出量分别为15.1±0.5(平均值±1标准误)和15.9±0.4毫升/分钟/100毫升(r = 0.96,p<0.01)。自动法和图形法测得的反流流量分别为2.1±0.2和2.3±0.2毫升/分钟/100毫升(r = 0.93,p<0.01)。这种体积描记仪可能是一种用于评估静脉疾病血流动力学改变的简单、快速且准确的设备。