Shellock F G, Shellock V J
Radnet and Future Diagnostics, Inc., Los Angeles, CA 90045, USA.
J Magn Reson Imaging. 1996 Nov-Dec;6(6):954-6. doi: 10.1002/jmri.1880060620.
The purpose of this study was to evaluate the MR compatibility of ceramic instruments. Nine different ceramic instruments were tested with respect to ferromagnetism, heating, and artifacts using previously described techniques. There was no magnetic field attraction, temperature increases were < or = 1 degree C, and artifacts involved a signal void similar to the size and shape of each instrument. The ceramic instruments were demonstrated to be compatible with MR imaging and acceptable for use during intraprocedural MR imaging.
本研究的目的是评估陶瓷器械的磁共振兼容性。使用先前描述的技术,对九种不同的陶瓷器械进行了铁磁性、加热和伪影方面的测试。没有磁场吸引,温度升高≤1℃,伪影表现为与每种器械的大小和形状相似的信号缺失。结果表明,这些陶瓷器械与磁共振成像兼容,可在术中磁共振成像期间使用。