Stone D L, Moheng M C, Rolih S, Sinor L T
Immucor, Inc., Norcross, Georgia 30071, USA.
J Clin Microbiol. 1997 Jan;35(1):217-22. doi: 10.1128/jcm.35.1.217-222.1997.
A solid-phase erythrocyte adherence assay has been developed for the serological detection of reagin antibodies in syphilis. Capture-S (Immucor, Inc., Norcross, Ga.) is a nontreponemal, qualitative screening test for the detection of immunoglobulin G (IgG) and IgM antilipid antibodies in serum or plasma samples from blood donors. The Capture-S assay utilizes a modified Venereal Disease Research Laboratory antigen bound to microtitration wells and anti-IgG- plus anti-IgM-coated indicator erythrocytes as the detection system. The Capture-S assay was evaluated at six separate sites on 10,942 specimens. For patient samples of clinically diagnosed syphilis categories (n = 366), the Capture-S assay yielded a sensitivity of 80.7% versus 80.3% for the rapid plasma reagin (RPR) card test (Becton Dickinson Microbiology Systems, Cockeysville, Md.). In comparative experiments on patient and donor samples (n = 10,222), the Capture-S assay demonstrated a sensitivity of 94% compared to 91.2% for the RPR card test. The Capture-S and RPR card tests produced essentially equivalent specificities of 99.2% and 99.3%, respectively, for this sample population. For five test sites, the Capture-S and RPR card test demonstrated a 98.3% agreement (10,085 of 10,264) of test results. These evaluations indicate that the Capture-S compares favorably to the RPR card test in assay sensitivity and specificity, with the added benefits of ease of use, accommodation of high-volume testing, and potential for automation.
已开发出一种固相红细胞黏附试验,用于梅毒中反应素抗体的血清学检测。Capture - S(Immucor公司,佐治亚州诺克罗斯)是一种非梅毒螺旋体定性筛查试验,用于检测献血者血清或血浆样本中的免疫球蛋白G(IgG)和IgM抗脂质抗体。Capture - S试验利用结合到微量滴定孔上的改良性病研究实验室抗原以及抗IgG和抗IgM包被的指示红细胞作为检测系统。在六个不同地点对10942份标本进行了Capture - S试验评估。对于临床诊断为梅毒的患者样本(n = 366),Capture - S试验的灵敏度为80.7%,而快速血浆反应素(RPR)卡片试验(Becton Dickinson微生物系统公司,马里兰州科基斯维尔)的灵敏度为80.3%。在对患者和献血者样本(n = 10222)的对比实验中,Capture - S试验的灵敏度为94%,而RPR卡片试验为91.2%。对于该样本群体,Capture - S试验和RPR卡片试验的特异性基本相当,分别为99.2%和99.3%。在五个测试地点,Capture - S试验和RPR卡片试验的检测结果一致性为98.3%(10264份中有10085份)。这些评估表明,Capture - S试验在检测灵敏度和特异性方面与RPR卡片试验相比具有优势,还具有使用方便、适合大量检测以及有自动化潜力等额外优点。