Macy E M, Hayes T E, Tracy R P
Department of Pathology, University of Vermont, Burlington 05405, USA.
Clin Chem. 1997 Jan;43(1):52-8.
We developed a reproducible ELISA for C-reactive protein (CRP), calibrated with WHO Reference Material, for which intra- and interassay CVs were 3.0% and 6.0%, respectively. Analytical recovery was 97.9%. The distribution of CRP in a healthy blood donor population (n = 143) was nongaussian, with 2.5th, 50th, and 97.5th percentile values of 0.08, 0.64, and 3.11 mg/L, respectively. There was no sex-related difference, and the association with age was weak. In a study of variability [by the method of Fraser and Harris (Crit Rev Clin Lab Sci 1989;27:409-37)], the analytical variability was 5.2%; the within-subject variability, CVI, was 42.2%; and the between-subject variability, CVG, was 92.5%. The critical difference for sequential values significant at P < or =0.05 (i.e., the smallest percentage change unlikely to be due to analytical variability or CVI) was calculated as 118%, and the index of individuality, CVI/CVG, was 0.46. This suggests that CRP, like many clinical chemistry analytes, has limited usefulness in detecting early disease-associated changes when used in conjunction with a healthy reference interval. From a molecular epidemiological standpoint, the usefulness of CRP in longitudinal studies is suggested by the small index of individuality and by observations that (a) short-term fluctuations were infrequent, (b) all data stayed within the reference interval, and (c) relative rankings of the subjects over 6 months only moderately deteriorated.
我们开发了一种用于检测C反应蛋白(CRP)的可重复性酶联免疫吸附测定(ELISA)方法,该方法采用世界卫生组织参考物质进行校准,其批内和批间变异系数分别为3.0%和6.0%。分析回收率为97.9%。143名健康献血者群体中CRP的分布呈非高斯分布,第2.5、50和97.5百分位数分别为0.08、0.64和3.11mg/L。未观察到性别差异,且与年龄的关联较弱。在一项变异性研究中[采用Fraser和Harris的方法(《临床实验室科学评论》1989年;27:409 - 37)],分析变异性为5.2%;个体内变异性CVI为42.2%;个体间变异性CVG为92.5%。计算得出在P≤0.05时具有显著性的连续值的临界差异(即不太可能由分析变异性或CVI导致的最小百分比变化)为118%,个体性指数CVI/CVG为0.46。这表明,与许多临床化学分析物一样,CRP与健康参考区间结合使用时,在检测早期疾病相关变化方面的作用有限。从分子流行病学角度来看,CRP在纵向研究中的作用体现在个体性指数较小以及以下观察结果:(a)短期波动不常见,(b)所有数据均保持在参考区间内,(c)6个月内受试者的相对排名仅略有下降。