[Probability of electrode dislocation: the importance of threshold measurements during pacemaker implantation (author's transl)].

作者信息

Rényi-Vámos F, Solti F, Gyöngy T, Szabó Z

出版信息

Z Kardiol. 1977 Jun;66(6):310-3.

PMID:899128
Abstract

Myocardial stimulation thresholds were measured during 774 pacemaker implantations with endocardial electrodes. It was found that myocardial thresholds were higher when using electrodes with greater surface area and bipolar ones. Early electrode dislocations were more frequent at higher myocardial thresholds. There was no relationship between frequency of late electrode dislocations and the value of myocardial thresholds. The importance of measurement of myocardial stimulation threshold - as a guarantee of successful pacemaker therapy - is emphasized.

摘要

在774例使用心内膜电极的起搏器植入过程中测量了心肌刺激阈值。发现使用表面积较大的电极和双极电极时心肌阈值较高。早期电极脱位在心肌阈值较高时更频繁。晚期电极脱位的频率与心肌阈值的值之间没有关系。强调了测量心肌刺激阈值作为起搏器治疗成功保证的重要性。