Suppr超能文献

一种用于临床硼中子俘获疗法中脑肿瘤照射的改进型中子准直器。

An improved neutron collimator for brain tumor irradiations in clinical boron neutron capture therapy.

作者信息

Liu H B, Greenberg D D, Capala J, Wheeler F J

机构信息

Medical Department, Brookhaven National Laboratory, Upton, New York 11973, USA.

出版信息

Med Phys. 1996 Dec;23(12):2051-60. doi: 10.1118/1.597772.

Abstract

To improve beam penetration into a head allowing the treatment of deeper seated tumors, two neutron collimators were built sequentially and tested for use in the clinical boron neutron capture therapy (BNCT) program at the epithermal neutron irradiation facility of the Brookhaven Medical Research Reactor. The collimators were constructed from lithium-impregnated polyethylene, which comprises Li2CO3 powder (approximately 93% enriched isotopic 6Li) uniformly dispersed in polyethylene to a total 6Li content of 7.0 wt. %. The first collimator is 7.6 cm thick with a conical cavity 16 cm in diameter on the reactor core side tapering to 8 cm facing the patient's head. The second collimator is 15.2 cm thick with a conical cavity 20 cm in diameter tapering to 12 cm. A clinical trial of BNCT for patients with malignant brain tumors is underway using the first collimator. Results of phantom dosimetry and Monte Carlo computations indicate that the new 15.2 cm thick collimator will improve the neutron beam penetration. Thus, the second collimator was made and will be used in an upcoming clinical trial. In-air and in-phantom mixed-field dosimetric measurements were compared to Monte Carlo computations for both collimators. The deeper penetration is achieved but at a sacrifice in beam intensity. In this report, a performance comparison of both collimators regarding various fluence rate and absorbed dose distributions in a head model is presented and discussed.

摘要

为了提高束流对头部的穿透能力,以便能够治疗更深部位的肿瘤,相继制造了两个中子准直器,并在布鲁克海文医学研究反应堆的超热中子辐照设施的临床硼中子俘获疗法(BNCT)项目中进行了测试。准直器由含锂聚乙烯制成,它由碳酸锂粉末(约93%富集的同位素6Li)均匀分散在聚乙烯中组成,锂的总含量为7.0 wt.%。第一个准直器厚7.6厘米,在反应堆堆芯一侧有一个直径16厘米的锥形腔,朝向患者头部一侧逐渐变细至8厘米。第二个准直器厚15.2厘米,有一个直径20厘米的锥形腔,逐渐变细至12厘米。目前正在使用第一个准直器对恶性脑肿瘤患者进行BNCT临床试验。模体剂量学和蒙特卡罗计算结果表明,新的15.2厘米厚的准直器将提高中子束的穿透能力。因此,制造了第二个准直器,并将用于即将进行的临床试验。将两个准直器的空气中和模体内混合场剂量测量结果与蒙特卡罗计算结果进行了比较。实现了更深的穿透,但牺牲了束流强度。在本报告中,对两个准直器在头部模型中的各种注量率和吸收剂量分布方面的性能进行了比较并展开了讨论。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验