Smith F P, Lora-Tamayo C, Carvajal R, Caddy B, Tagliaro F
Graduate Program in Forensic Science, University of Alabama at Birmingham, USA.
Ann Clin Biochem. 1997 Jan;34 ( Pt 1):81-4. doi: 10.1177/000456329703400112.
A recently introduced automated immunoassay which is based on kinetic interaction of microparticles in solution (Roche ONLINE), was evaluated for the detection of cocaine metabolite benzoylecgonine (BE) and opiates in human urine. Cross-reactivity for the opiates morphine (100%), codeine (88%), 6-monoacetylmorphine (88%), and morphine 3-glucuronide (72%) was assessed. Analytical recovery evaluated on blank urines spiked with 0, 250, 300, 350, and 500 micrograms/L of morphine and BE (n = 10), varied from 85.2 to 100.2% for opiates and from 81.4 to 93.1% for the cocaine metabolite. The within-day precision ranged from 1.4 to 4.7% for morphine and from 4.2 to 4.8% for BE. The repeatability of the standards over 1 month was 1.0-3.3% for opiates and 1.7-5.1% for BE, and thus allowing measurements to continue over 30 days without re-calibration. This method compared favourably with the SYVA EMIT d.a.u system and gas chromatography/mass spectroscopy (GC/MS) methods.