Zhu Y, Kirov A S, Mishra V, Meigooni A S, Williamson J F
Radiation Oncology Center, Washington University School of Medicine, St. Louis, Missouri 63110, USA.
Med Phys. 1997 Feb;24(2):223-31. doi: 10.1118/1.598068.
Radiochromic film (RCF) is attractive as a thin, high resolution, 2D planar dosimeter. We have studied the uniformity, linearity, and reproducibility of a commercially supplied RCF system (model MD-55). Forty 12 cm long strips of RCF were exposed to uniform doses of 6 MV x rays. Optical density (OD) distributions were measured by a helium-neon scanning laser (633 nm) 2D densitometer and also with a manual densitometer. All film strips showed 8%-15% variations in OD values independent of densitometry technique which are evidently due to nonuniform dispersal of the sensor medium. A double exposure technique was developed to solve this problem. The film is first exposed to a uniform beam, which defines a pixel-by-pixel nonuniformity correction matrix. The film is then exposed to the unknown dose distribution, rescanned, and the net OD at each pixel corrected for nonuniformity. The double exposure technique reduces OD/unit dose variation to a 2%-5% random fluctuation. RCF response was found to deviate significantly from linearity at low doses (40% change in net OD/Gy from 1 to 30 Gy); a finding not previously reported. To study the tradeoff between statistical noise and spatial resolution, OD was averaged over blocks of adjacent 50 microns pixels (ranging from 1 x 1 to 10 x 10 pixels). Reproducibility, defined as the standard deviation of repeated single-pixel measurements on separate film pieces, was 2% at 30 Gy for a resolution of 0.25 mm. With careful correction for nonlinearity and nonuniformity, RCF is a promising quantitative 2D dosimeter for radiation oncology applications.
放射变色薄膜(RCF)作为一种薄型、高分辨率的二维平面剂量计很有吸引力。我们研究了一种商业供应的RCF系统(型号MD - 55)的均匀性、线性和可重复性。四十条12厘米长的RCF条带被暴露于均匀剂量的6兆伏X射线。通过氦氖扫描激光(633纳米)二维密度计以及手动密度计测量光密度(OD)分布。所有薄膜条带的OD值显示出8% - 15%的变化,与密度测量技术无关,这显然是由于传感器介质的不均匀分散所致。开发了一种双曝光技术来解决这个问题。薄膜首先暴露于均匀光束下,这定义了一个逐像素的不均匀性校正矩阵。然后薄膜暴露于未知剂量分布,重新扫描,并对每个像素的净OD进行不均匀性校正。双曝光技术将OD/单位剂量变化降低到2% - 5%的随机波动。发现RCF响应在低剂量时显著偏离线性(从1到30戈瑞,净OD/戈瑞变化40%);这一发现此前未被报道。为了研究统计噪声和空间分辨率之间的权衡,对相邻50微米像素块(范围从1×1到10×10像素)的OD进行平均。可重复性定义为在单独薄膜片上重复单像素测量的标准偏差,对于0.25毫米的分辨率,在30戈瑞时为2%。经过仔细的非线性和不均匀性校正后,RCF是一种用于放射肿瘤学应用的有前景的定量二维剂量计。