Busch M, Makoski B, Schulz U, Sauerwein K
Strahlentherapie. 1977 Sep;153(9):581-8.
An afterloading system for the intracavitary radiotherapy is described which works with a unique radiation source of 10 Ci 192Ir moving in steps of 1 cm. The times during which the source stops can be determined at a control panel. By using an intermediate piece of 10 cm, a total course of the source of 20 cm can be obtained. The course of the source can be calculated individually by a computer program. This program even takes into consideration an additional deep therapy if the position of the deep therapy field in the target volume is known. This method allows to adapt the total isodoses very precisely to the course of the tolerance doses.
描述了一种用于腔内放射治疗的后装系统,该系统使用独特的10居里铱-192辐射源,以1厘米的步长移动。源停止的时间可在控制面板上确定。通过使用10厘米的中间部件,可获得源的20厘米总行程。源的行程可通过计算机程序单独计算。如果已知深部治疗野在靶体积中的位置,该程序甚至会考虑额外的深部治疗。这种方法能够非常精确地使总等剂量与耐受剂量的过程相匹配。