Pincivero D M, Lephart S M, Karunakara R A
Neuromuscular Research Laboratory, University of Pittsburgh, PA, USA.
Int J Sports Med. 1997 Feb;18(2):113-7. doi: 10.1055/s-2007-972605.
The importance of accurately reproducing isokinetic strength values is critical for the assessment of training induced changes in muscle function. The purpose of this study was to determine the test re-test reliability of the Biodex System 2 Isokinetic Dynamometer for concentric quadricep and hamstring strength and muscular endurance. Twenty-one healthy volunteers underwent isokinetic testing on 2 occasions separated by 7 days. Each subject performed 5 repetitions at 60 degrees/ second followed by a muscular endurance test which consisted of 30 repetitions at 180 degrees/second. The results demonstrated high ICC values ranging from r = 0.88 to r = 0.97 and r = 0.82 to r = 0.96 for variables measured at 60 degrees/second and 180 degrees/second, respectively, SEM values were found to range from low to moderate, 4.8% to 11.6% and 5.6% to 10.8%, at 60 and 180 degrees/second, respectively. ICC coefficients were found to be low, r = 0.52 to r = 0.74, and SEM values were found to be high, 9.8% to 20.8%, for the work fatigue index at 180 degrees/second. The results demonstrate that isokinetic values obtained at 60 and 180 degrees/second are highly reproducible with the Biodex System 2 Dynamometer provided that adequate calibration, gravity correction and patient positioning is recorded and standardized. However, the work fatigue index as a measure of muscular endurance has not been shown to be reliable or precise. Changes in muscular strength over time should be considered valid if these differences lie beyond the SEM values reported in this study.
准确再现等速肌力值对于评估训练引起的肌肉功能变化至关重要。本研究的目的是确定Biodex System 2等速测力计对股四头肌和腘绳肌向心收缩力量及肌肉耐力的重测信度。21名健康志愿者分两次进行等速测试,间隔7天。每位受试者在60度/秒的速度下进行5次重复测试,随后进行肌肉耐力测试,该测试包括在180度/秒的速度下进行30次重复测试。结果表明,在60度/秒和180度/秒测量的变量的组内相关系数(ICC)值较高,分别为r = 0.88至r = 0.97和r = 0.82至r = 0.96,标准误(SEM)值在60度/秒和180度/秒时分别为低到中度,即4.8%至11.6%和5.6%至10.8%。对于180度/秒时的工作疲劳指数,ICC系数较低,r = 0.52至r = 0.74,SEM值较高,为9.8%至20.8%。结果表明,只要记录并标准化了适当的校准、重力校正和患者体位,使用Biodex System 2测力计在60度/秒和180度/秒时获得的等速值具有高度可重复性。然而,作为肌肉耐力指标的工作疲劳指数尚未被证明是可靠或精确的。如果肌肉力量随时间的变化超出本研究报告的SEM值,则应认为这些变化是有效的。