健康受试者视网膜厚度的体内定量测量。

Quantitative in vivo retinal thickness measurements in healthy subjects.

作者信息

Landau D, Schneidman E M, Jacobovitz T, Rozenman Y

机构信息

The Ophthalmology Department, Shaare Zedek Medical Center, Jerusalem, Israel.

出版信息

Ophthalmology. 1997 Apr;104(4):639-42. doi: 10.1016/s0161-6420(97)30259-0.

Abstract

PURPOSE

The early detection of visual threatening retinal thickness alterations is important for the purpose of offering affected patients proven treatment when indicated. Currently, the clinical methods that are available for obtaining an impression of retinal thickness are subjective and require experience and expertise. The authors present measurements of foveal thickness in healthy individuals obtained by a newly developed instrument that enables a noninvasive, noncontact, safe, and fast measurement of the retinal thickness anywhere in the posterior pole.

METHODS

A prototype of the retinal thickness analyzer (RTA) that operates on the principle of laser-slit biomicroscopy was used. Retinal thickness was measured in healthy emmetropic volunteers.

RESULTS

Fifty eyes were measured. The average thickness at the foveola was 178 +/- 44 microm (+/- standard deviation; range, 100-322 microm).

CONCLUSION

The retinal thickness measured by the RTA in healthy subjects correspond well to previous published data (in vivo and histologic) on retinal thickness. This instrument may prove valuable in detecting retinal thickness alteration in macular diseases.

摘要

目的

尽早发现有视力威胁的视网膜厚度改变对于在必要时为受影响患者提供经证实的治疗非常重要。目前,用于获取视网膜厚度印象的临床方法具有主观性,需要经验和专业知识。作者展示了通过一种新开发的仪器获得的健康个体中央凹厚度测量值,该仪器能够对后极部任何部位的视网膜厚度进行无创、非接触、安全且快速的测量。

方法

使用了基于激光裂隙生物显微镜原理运行的视网膜厚度分析仪(RTA)原型。对健康正视志愿者进行视网膜厚度测量。

结果

测量了50只眼睛。小凹处的平均厚度为178±44微米(±标准差;范围为100 - 322微米)。

结论

RTA在健康受试者中测量的视网膜厚度与先前发表的关于视网膜厚度的数据(体内和组织学数据)非常吻合。该仪器在检测黄斑疾病中的视网膜厚度改变方面可能具有重要价值。