Suppr超能文献

通过眼底监测记录多焦视网膜电图。

Recording multifocal electroretinograms with fundus monitoring.

作者信息

Kondo M, Miyake Y, Horiguchi M, Suzuki S, Tanikawa A

机构信息

Department of Ophthalmology, Nagoya University School of Medicine, Japan.

出版信息

Invest Ophthalmol Vis Sci. 1997 Apr;38(5):1049-52.

PMID:9113002
Abstract

PURPOSE

To record multifocal electroretinograms (ERGs) with simultaneous fundus monitoring.

METHODS

An infrared television fundus camera was used to monitor the fundus. A tightly packed array of 19 yellow light-emitting diodes (LEDs) was used as the stimulus source and LEDs were alternated between on and off according to a binary m-sequence at a rate of 75 Hz. The stimulus array subtended approximately 25 degrees of the visual angle. Multifocal ERGs were recorded with a Burian-Allen bipolar contact lens electrode from two normal subjects and a patient with macular dystrophy.

RESULTS

When the center of the stimulus was positioned on the fovea, the amplitude of the response was largest at the fovea and decreased toward the peripheral retina; the response pattern corresponded to the spatial distribution of cone cells. When the center of the stimulus was positioned on the optic disc, the response at the optic disc was undetectable. A reliable multifocal ERG was obtained from a patient with macular dystrophy.

CONCLUSIONS

This system made it possible to monitor the stimulus on the fundus and to adjust the stimulus to the correct retinal locations with exact focusing. The authors' findings indicate that this method especially can be useful in patients with poor fixation. In addition, the authors' results suggests that the stray light effect in the multifocal stimulus system is minimal if any.

摘要

目的

记录多焦视网膜电图(ERG)并同时进行眼底监测。

方法

使用红外电视眼底相机监测眼底。一个紧密排列的由19个黄色发光二极管(LED)组成的阵列用作刺激源,LED根据二进制m序列以75Hz的速率在开和关之间交替。刺激阵列所对视角约为25度。用Burian-Allen双极接触镜电极记录两名正常受试者和一名黄斑营养不良患者的多焦ERG。

结果

当刺激中心位于中央凹时,中央凹处的反应幅度最大,并向周边视网膜减小;反应模式与视锥细胞的空间分布相对应。当刺激中心位于视盘时,视盘处的反应无法检测到。从一名黄斑营养不良患者获得了可靠的多焦ERG。

结论

该系统能够监测眼底的刺激,并通过精确聚焦将刺激调整到正确的视网膜位置。作者的研究结果表明,这种方法对注视不良的患者尤其有用。此外,作者的结果表明,多焦刺激系统中的杂散光效应即使有也是最小的。