Respess R A, Butcher A, Wang H, Chaowanachan T, Young N, Shaffer N, Mastro T D, Biryahwaho B, Downing R, Tanuri A, Schechter M, Pascu R, Zekeng L, Kaptué L, Gürtler L, Eberle J, Ellenberger D, Fridlund C, Rayfield M, Kwok S
Roche Molecular Systems, Inc., Alameda, California 94501, USA.
J Clin Microbiol. 1997 May;35(5):1284-6. doi: 10.1128/jcm.35.5.1284-1286.1997.
A panel of 136 genetically diverse group M and 5 group O adult isolates from outside the United States and Europe were evaluated by PCR with the Roche AMPLICOR HIV-1 test, a modified version of the AMPLICOR HIV-1 test, and a new primer pair/probe system. Detection of some of these isolates was less efficient with the AMPLICOR HIV-1 test; however, the assay was significantly improved by reducing the sample input and lowering the annealing temperature. The new primer pair/probe set detected 140 of 141 isolates, including the 5 group O isolates that were not detected with either of the AMPLICOR HIV-1 test formats.