泰尔茂Capiox离心泵搏动流模式的临床评估。

Clinical evaluation of pulsatile flow mode of Terumo Capiox centrifugal pump.

作者信息

Nishida H, Uesugi H, Nishinaka T, Uwabe K, Aomi S, Endo M, Koyanagi H, Oshiyama H, Nogawa A, Akutsu T

机构信息

Department of Cardiovascular Surgery, Heart Institute of Japan, Tokyo Women's Medical College, Japan.

出版信息

Artif Organs. 1997 Jul;21(7):816-21. doi: 10.1111/j.1525-1594.1997.tb03749.x.

Abstract

The Terumo Capiox centrifugal pump system possesses an automatic priming function in which the motor repeatedly stops and runs intermittently to eliminate air bubbles in the circuit through the micropores of the hollow-fiber membrane oxygenator. By modifying this mechanism, we have developed a pulsatile flow mode. In this mode, maximum and minimum pump rotational speeds can be independently set every 20 rpm in the range of 0 to 3,000 rpm. The duration of the pump run at maximum and minimum speeds can also be independently set every 0.1 s in the range of 0.2 to 15 s. In a clinical trial, after obtaining the desired flow rate, 2.4 L/min/m2 in nonpulsatile flow mode, a pulsatile flow mode of 60 cycles/min (with 1 cycle being maximum speed for 0.4 s and minimum speed for 0.6 s) was obtained by adding and subtracting 500 rpm to and from the rotational speed in nonpulsatile flow mode. Pulse pressures in the femoral artery and in the circuit just proximal to the perfusion cannula (6.5 mm Sarns high flow cannula with metal tip) were measured in 5 patients who underwent pulsatile cardiopulmonary bypass (CPB) for a coronary artery bypass graft (CABG), and compared to pulse pressures obtained by intraaortic balloon pumping (IABP) in 3 patients and by the pulsatile mode of the 3M Delphin pump in 3 patients. The platelet count, free hemoglobin, and beta-thromboglobulin (beta-TG) were measured and compared with measurements from another 5 patients who underwent nonpulsatile CPB. Although the pulse pressure measured in the circuit was 180 mm Hg on average, the pressure in the femoral artery was only 15 to 40 mm Hg with a mean of 20 mm Hg. In the same patients, 60 to 80 mm Hg pulse pressure was obtained with IABP. The pulse pressure obtained with the Delphin pump was not more than that obtained with the Terumo pump. There were no significant differences in percents of preoperative levels of platelet counts (pulsatile, 87.6 +/- 15.8% and nonpulsatile, 72.4 +/- 40.6%), free hemoglobin (pulsatile, 18 +/- 8 mg/dl and nonpulsatile, 25 = 7 mg/dl), and beta-TG (pulsatile 298 +/- 28 ng/ml and nonpulsatile, 312 +/- 143 ng/ml). In conclusion, although the pulsatile mode of the Terumo centrifugal pump did not exhibit any adverse effects hematologically, the pulse pressure obtained was unsatisfactorily small, mainly because of dumping caused by the perfusion cannula.

摘要

泰尔茂Capiox离心泵系统具有自动充液功能,即电机反复停止和间歇运行,以通过中空纤维膜氧合器的微孔消除回路中的气泡。通过改进这一机制,我们开发了一种搏动血流模式。在这种模式下,泵的最大和最小转速可在0至3000转/分钟的范围内以每20转/分钟的间隔独立设置。泵在最大和最小速度下运行的持续时间也可在0.2至15秒的范围内以每0.1秒的间隔独立设置。在一项临床试验中,在获得所需流量(非搏动血流模式下为2.4升/分钟/平方米)后,通过在非搏动血流模式下的转速上加、减500转/分钟,获得了60次/分钟的搏动血流模式(1个周期为最大速度0.4秒,最小速度0.6秒)。对5例行冠状动脉旁路移植术(CABG)的搏动性体外循环(CPB)患者的股动脉和灌注插管近端(6.5毫米Sarns高流量金属头插管)回路中的脉压进行了测量,并与3例行主动脉内球囊反搏(IABP)患者和3例行3M德尔芬泵搏动模式患者的脉压进行了比较。测量了血小板计数、游离血红蛋白和β-血小板球蛋白(β-TG),并与另外5例行非搏动性CPB患者的测量结果进行了比较。尽管回路中测量的脉压平均为180毫米汞柱,但股动脉中的压力仅为15至

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验