Peele J D, Gadsden R H, Crews R
Clin Chem. 1977 Dec;23(12):2238-41.
Reproducibility of reading "N-Multistix" dipsticks by a semi-automated urinalysis instrument (Ames' "Clini-Tek") has been described for artifically prepared samples. Glucose, ketone, urobilinogen, and nitrite showed high reproducibility (greater than 90%) for reading multiple samples at predetermined analyte concentrations. Determination of proteinuria showed the lowest proportion of false positives (2-3%) and false negatives (0%). Determination of hemoglobinuria and bilirubinuria by dipsticks were the least reproducible. Urobilinogen showed no interference from bilirubin in concentrations up to 32 mg/liter. Precision was high for results for quality-control capsules provided by the manufacturer.
已描述了半自动尿液分析仪(艾姆斯公司的“Clini-Tek”)读取“N-Multistix”试纸条对人工制备样本的重复性。对于在预定分析物浓度下读取多个样本,葡萄糖、酮体、尿胆原和亚硝酸盐显示出高重复性(大于90%)。蛋白尿测定显示假阳性(2 - 3%)和假阴性(0%)的比例最低。试纸条检测血红蛋白尿和胆红素尿的重复性最差。尿胆原在浓度高达32毫克/升时不受胆红素干扰。制造商提供的质量控制胶囊结果的精密度很高。