Cheng C W, Das I J, Tang W, Chang S, Tsai J S, Ceberg C, De Gaspie B, Singh R, Fein D A, Fowble B
Department of Radiation Oncology, University of Arizona, Tucson 85724-5081, USA.
Int J Radiat Oncol Biol Phys. 1997 Jul 1;38(4):835-42. doi: 10.1016/s0360-3016(97)00078-3.
The objectives of this study are: (1) to investigate the dosimetric differences of the different treatment planning systems (TPS) in breast irradiation with tangential fields, and (2) to study the effect of beam characteristics on dose distributions in tangential breast irradiation with 6 MV linear accelerators from different manufacturers.
Nine commercial and two university-based TPS are evaluated in this study. The computed tomographic scan of three representative patients, labeled as "small", "medium" and "large" based on their respective chest wall separations in the central axis plane (CAX) were used. For each patient, the tangential fields were set up in each TPS. The CAX distribution was optimized separately with lung correction, for each TPS based on the same set of optimization conditions. The isodose distributions in two other off-axis planes, one 6 cm cephalic and the other 6 cm caudal to the CAX plane were also computed. To investigate the effect of beam characteristics on dose distributions, a three-dimensional TPS was used to calculate the isodose distributions for three different linear accelerators, the Varian Clinac 6/100, the Siemens MD2 and the Philips SL/7 for the three patients. In addition, dose distributions obtained with 6 MV X-rays from two different accelerators, the Varian Clinac 6/100 and the Varian 2100C, were compared.
For all TPS, the dose distributions in all three planes agreed qualitatively to within +/- 5% for the "small" and the "medium" patients. For the "large" patient, all TPS agreed to within +/- 4% on the CAX plane. The isodose distributions in the caudal plane differed by +/- 5% among all TPS. In the cephalic plane in which the patient separation is much larger than that in the CAX plane, six TPS correctly calculated the dose distribution showing a cold spot in the center of the breast contour. The other five TPS showed that the center of the breast received adequate dose. Isodose distributions for 6 MV X-rays from three different accelerators differed by about +/- 3% for the "small" patient and more than +/- 5% for the "large" patient. For two different 6 MV machines of the same manufacturer, the isodose distribution agreed to within +/- 2% for all three planes for the "large" patient.
The differences observed among the various TPS in this study were within +/- 5% for both the "small" and the "medium" patients while doses at the hot spot exhibit a larger variation. The large discrepancy observed in the off-axis plane for the "large" patient is largely due to the inability of most TPS to incorporate the collimator angles in the dose calculation. Only six systems involved agreed to within +/- 5% for all three patients in all calculation planes. The difference in dose distributions obtained with three accelerators from different manufacturers is probably due to the difference in beam profiles. On the other hand, the 6 MV X-rays from two different models of linear accelerators from the same manufacturer have similar beam characteristics and the dose distributions are within +/- 2% of each other throughout the breast volume. In general, multi-institutional breast treatment data can be compared within a +/- 5% accuracy.
本研究的目标为:(1)研究不同治疗计划系统(TPS)在乳腺切线野照射中的剂量学差异;(2)研究不同厂家的6兆伏直线加速器的射束特性对乳腺切线野照射剂量分布的影响。
本研究评估了九种商用和两种基于大学的TPS。使用了三名代表性患者的计算机断层扫描,根据其在中心轴平面(CAX)上各自的胸壁间距分别标记为“小”、“中”和“大”。对于每位患者,在每个TPS中设置切线野。基于同一组优化条件,在每个TPS中分别对CAX分布进行肺部校正优化。还计算了另外两个离轴平面中的等剂量分布,一个位于CAX平面上方6厘米,另一个位于CAX平面下方6厘米。为了研究射束特性对剂量分布的影响,使用三维TPS计算了三名患者使用三种不同直线加速器(瓦里安Clinac 6/100、西门子MD2和飞利浦SL/7)的等剂量分布。此外,比较了来自两种不同加速器(瓦里安Clinac 6/100和瓦里安2100C)的6兆伏X射线获得的剂量分布。
对于所有TPS,“小”和“中”患者在所有三个平面中的剂量分布在定性上一致,偏差在±5%以内。对于“大”患者,所有TPS在CAX平面上的偏差在±4%以内。所有TPS在尾侧平面中的等剂量分布相差±5%。在头侧平面中,患者间距比CAX平面中的大得多,六个TPS正确计算出剂量分布,显示乳腺轮廓中心有冷点。其他五个TPS显示乳腺中心接受了足够的剂量。对于“小”患者,三种不同加速器的6兆伏X射线的等剂量分布相差约±3%,对于“大”患者相差超过±5%。对于同一厂家的两种不同6兆伏机器,“大"患者在所有三个平面中的等剂量分布偏差在±2%以内。
本研究中观察到的不同TPS之间的差异,对于“小”和“中”患者在±5%以内,而热点处的剂量变化较大。在“大”患者的离轴平面中观察到的较大差异主要是由于大多数TPS无法在剂量计算中纳入准直器角度。只有六个系统在所有计算平面中对所有三名患者的偏差在±5%以内。不同厂家的三种加速器获得的剂量分布差异可能是由于射束轮廓的差异。另一方面,来自同一厂家的两种不同型号直线加速器的6兆伏X射线具有相似的射束特性,并且在整个乳腺体积中的剂量分布彼此相差在±2%以内。一般来说,多机构乳腺治疗数据可以在±5%的精度内进行比较。