Survival for up to six months in calves supported with an implantable axial flow ventricular assist device.

作者信息

Macha M, Litwak P, Yamazaki K, Kameneva M, Kerrigan J P, Antaki J F, Butler K C, Thomas D C, Taylor L P, Kormos R L, Griffith B P, Borovetz H S

机构信息

McGowan Center, Department of Surgery, University of Pittsburgh Medical Center, Pennsylvania 15219, USA.

出版信息

ASAIO J. 1997 Jul-Aug;43(4):311-5.

PMID:9242945
Abstract

This paper summarizes the authors' in vivo experience to date with an implantable axial flow blood pump designed for long-term ventricular support. This small, valveless pump with blood-lubricated bearings has been implanted in six calves (83 +/- 6 kg) as a left ventricular assist device (LVAS). The left ventricle and descending thoracic aorta were cannulated by left thoracotomy, and the pump was placed in a subcutaneous pocket below the costal margin. Animals remained hemodynamically stable throughout the course of support during partial left ventricular bypass. Five animals were killed after 15, 27, 52, 57, and 181 days. The longest survivor (181 days) demonstrated normal pump function at the time death. Pump speed was maintained at 10,100 +/- 100 rpm, with an average pump flow rate of 4.9 +/- 0.5 L/min under resting physiologic conditions. Average plasma free hemoglobin was 17.4 +/- 7.5 mg/dl. Renal, hepatic, and hematologic indices remained within physiologic range in all of these animals, except during the immediate postoperative period. Histopathologic analyses of major organs after death revealed small renal cortical infarcts in five of six animals; the remaining organs were normal. These animal studies support the feasibility of this small implanted axial flow pump for long-term ventricular assistance.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验