无针接头导致导管微生物污染的可能性。

The potential for catheter microbial contamination from a needleless connector.

作者信息

Brown J D, Moss H A, Elliott T S

机构信息

Department of Clinical Microbiology, Queen Elizabeth Hospital, Edgbaston, Birmingham, UK.

出版信息

J Hosp Infect. 1997 Jul;36(3):181-9. doi: 10.1016/s0195-6701(97)90193-6.

Abstract

Needleless connectors have been widely introduced into clinical practice to allow the connection of syringes and luers to peripheral and central vascular catheters. The potential for microbial contamination of catheters via these devices is currently unclear. A recently introduced connector, the 'Connecta Clave', was assessed by various in-vitro methods. The 'Connecta Clave' is specifically devised to separate external components from the fluid pathway. The compression seals of 50 devices were contaminated with 1 x 10(4) cfu Staphylococcus epidermidis, disinfected with isopropanol, and fluid passed through. Only one device allowed organisms to pass through, despite this challenge, representing a contamination rate of 2%. In comparison, when 50 connectors were challenged with 20 cfu of S. epidermidis, no organisms passed through the device during use. In the clinical situation, after manipulation, < 16 cfu of skin organisms were found associated with the compression seal of the devices. It is, therefore, likely that the contamination rates in clinical practice will be extremely low. Three methods of disinfecting the compression seals and associated rims were also evaluated. A combination of alcohol chlorhexidine spray, followed by a 70% isopropanol swab, resulted in the most efficacious disinfection. The isopropanol swabs produced an adequate disinfection rate. The overall results suggest that by use of specially designed connectors, not only are needlestick injuries reduced, but the likelihood of microbial contamination of catheters via the internal route may also be diminished.

摘要

无针连接器已广泛应用于临床实践,用于将注射器和鲁尔接头连接到外周和中心血管导管。目前尚不清楚通过这些装置导致导管微生物污染的可能性。一种最近推出的连接器“Connecta Clave”通过各种体外方法进行了评估。“Connecta Clave”专门设计用于将外部组件与流体通路分开。50个装置的压缩密封件被1×10⁴ cfu表皮葡萄球菌污染,用异丙醇消毒,然后使流体通过。尽管面临这种挑战,但只有一个装置允许微生物通过,污染率为2%。相比之下,当50个连接器受到20 cfu表皮葡萄球菌的挑战时,使用过程中没有微生物通过该装置。在临床情况下,操作后,发现与装置压缩密封件相关的皮肤微生物<16 cfu。因此,临床实践中的污染率可能极低。还评估了三种对压缩密封件和相关边缘进行消毒的方法。酒精洗必泰喷雾,然后用70%异丙醇棉签擦拭,可产生最有效的消毒效果。异丙醇棉签产生了足够的消毒率。总体结果表明,通过使用专门设计的连接器,不仅可以减少针刺伤,而且通过内部途径导致导管微生物污染的可能性也可能降低。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验