Can J Surg. 1997 Aug;40(4):310-2.
The smaller acetabular components used in total hip replacement may have a polyethylene liner that is too thin, resulting in higher polyethylene stress and an increased potential for wear. The authors present a case that highlights the problem of acute polyethylene fracture. To compensate for the thinness of the polyethylene, the authors recommend the use of a smaller head size to allow polyethylene thickness of at least 8 mm.