Department of Orthopaedic Surgery, Somdej Prapinkao Hospital, Bangkok, Thailand.
J Arthroplasty. 2012 Feb;27(2):323.e1-4. doi: 10.1016/j.arth.2011.04.010. Epub 2011 May 20.
Highly cross-linked polyethylene (HXPE) in total hip arthroplasty has been shown to decrease wear rate compared with conventional liner. However, it has some disadvantages in that the mechanical properties cause early failure of the implant. This case report presents an unexpected failure of total hip arthroplasty in a 72-year-old woman that occurred at 20 months postsurgery. Operative findings revealed fracture of superior rim at locking groove of liner. We concluded that the failure was caused by decreased mechanical properties of highly cross-linked polyethylene, less thickness of polyethylene, more vertical cup, and use of large femoral head.
全髋关节置换术中的超高分子量聚乙烯(HXPE)与传统衬垫相比,已被证明可降低磨损率。然而,它也存在一些缺点,即其机械性能会导致植入物早期失效。本病例报告介绍了一位 72 岁女性在术后 20 个月时发生的全髋关节置换术意外失败。手术发现显示衬垫锁定槽的上缘骨折。我们得出结论,失败的原因是超高分子量聚乙烯的机械性能降低、聚乙烯厚度变薄、杯体更垂直以及使用大尺寸股骨头。