Early failure of the femoral component in unicompartmental knee arthroplasty.

作者信息

Riebel G D, Werner F W, Ayers D C, Bromka J, Murray D G

机构信息

Schenectady Orthopaedic Associates, PC, Schenectady, New York 12308, USA.

出版信息

J Arthroplasty. 1995 Oct;10(5):615-21. doi: 10.1016/s0883-5403(05)80205-0.

Abstract

Unicompartmental knee arthroplasty has been proposed as treatment for unicompartmental arthritis. Overall, results have been satisfactory, with occasional failures of tibial component fixation. Short-term loosening of the femoral components in unicompartmental knee arthroplasty that required revision to total knee arthroplasty is reported. All patients presented with pain and instability of the knee, with loosening confirmed radiographically and at revision surgery. The Porous-Coated Anatomic femoral component (Howmedica, Rutherford, NJ) was biomechanically tested after being implanted in cadaver limbs. Cyclic loading in flexion and extension under physiologic loads resulted in implant failure. The mode of failure was rocking in the sagittal plane, which corresponds to the clinical findings. The femoral implant design consists of an angled bone-implant interface with straight arms, which causes shear at the bone-prosthesis interface did not loosen under identical biomechanical testing in paired knees. It was concluded that the Porous-Coated Anatomic femoral component has an unacceptably high early clinical loosening rate, confirmed by comparative biomechanical testing.

摘要

单髁膝关节置换术已被提议用于治疗单髁关节炎。总体而言,结果令人满意,但偶尔会出现胫骨组件固定失败的情况。有报道称,单髁膝关节置换术中股骨组件短期松动,需要翻修为全膝关节置换术。所有患者均表现为膝关节疼痛和不稳定,影像学检查及翻修手术证实存在松动。多孔涂层解剖型股骨组件(Howmedica,新泽西州卢瑟福)植入尸体肢体后进行了生物力学测试。在生理负荷下进行屈伸循环加载导致植入物失败。失败模式为矢状面摇摆,这与临床发现相符。股骨植入物设计包括一个带有直臂的成角骨-植入物界面,在配对膝关节的相同生物力学测试下,该界面在骨-假体界面处产生剪切力,但并未松动。得出的结论是,通过比较生物力学测试证实,多孔涂层解剖型股骨组件的早期临床松动率高得令人无法接受。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验