Halpren B A, Axline S G, Coplon N S, Brown D M
J Infect Dis. 1976 Jun;133(6):627-36. doi: 10.1093/infdis/133.6.627.
The effects of four types of artificial kidney on dialyzer clearance rates and serum pharmacokinetics of gentamicin were compared. In 19 patients undergoing chronic hemodialysis, the mean (+/-SE) interdialysis half-life of gentamicin in serum was 49.3 +/- 3.5 hr, whereas during dialysis this value was reduced to 10.0 +/- 0.7 hr. The mean half-life of gentamicin in serum at conventional flow rates for the Hollow Fiber Kidney, Kiil, Gambro, and Coil dialyzers was 11.3, 10.9, 8.2, and 7.4 hr, respectively, and the mean values for clearance of gentamicin were 26,28,42, and 48 ml per min, respectively. For all dialyzers, rates of clearance of gentamicin increased linearly with plasma flow rate over the flow range studied. The Gambro and Coil dialyzers had significantly higher rates of clearance of gentamicin from serum (P less than 0.05) than the Hollow Fiber Kidney and Kiil dialyzers over a wide range of clinically useful plasma flow rates (119-300 ml per min), whereas the Kiil dialyzer cleared gentamicin more effectively (P less than 0.05) than the Hollow Fiber Kidney dialyzer over a more limited interval (117-177 ml per min). Therapeutic recommendations for patients undergoing hemodialysis were made in light of current findings.
比较了四种类型的人工肾对透析器清除率和庆大霉素血清药代动力学的影响。在19例接受慢性血液透析的患者中,血清中庆大霉素透析间期的平均(±标准误)半衰期为49.3±3.5小时,而透析期间该值降至10.0±0.7小时。中空纤维肾、基尔型、金宝型和盘管型透析器在常规流速下血清中庆大霉素的平均半衰期分别为11.3、10.9、8.2和7.4小时,庆大霉素的平均清除率分别为每分钟26、28、42和48毫升。对于所有透析器,在所研究的流速范围内,庆大霉素的清除率随血浆流速呈线性增加。在广泛的临床实用血浆流速(每分钟119 - 300毫升)范围内,金宝型和盘管型透析器从血清中清除庆大霉素的速率显著高于中空纤维肾和基尔型透析器(P小于0.05),而在更有限的区间(每分钟117 - 177毫升)内,基尔型透析器清除庆大霉素比中空纤维肾透析器更有效(P小于0.05)。根据当前研究结果,对接受血液透析的患者提出了治疗建议。