Prestele K, Franetzki M, Meerlender G
Respiration. 1976;33(2):150-62. doi: 10.1159/000193728.
For the estimation and correction of errors in the measurement of the respiratory flow with viscosity-affected respiratory flow receptors the viscosity of natural respiratory gases (N2, O2, CO2, H2O) and of the ternary system helium-oxygen-nitrogen, which is of importance in pulmonary function diagnosis, was determined. Corresponding density values were calculated. Allowance was made for such special features of spirometry as exchange of O2 for CO2 and water vapour saturation at 37 degrees C. The viscosity of the ternary system is shown in a graph. The formulae indicated have been simplified as far as this was compatible with the demands of accuracy. The fluctuations in density and viscosity of natural respiratory gases can be read directly from a table.
为了用受粘度影响的呼吸流量感受器估算和校正呼吸流量测量中的误差,测定了天然呼吸气体(N₂、O₂、CO₂、H₂O)以及在肺功能诊断中具有重要意义的三元系统氦 - 氧 - 氮的粘度。计算了相应的密度值。考虑到肺量计的一些特殊特性,如O₂与CO₂的交换以及37℃时的水蒸气饱和。三元系统的粘度以图表形式给出。所给出的公式已在符合精度要求的前提下尽可能简化。天然呼吸气体密度和粘度的波动可直接从表格中读取。