核磁共振相控阵幅度图像中的信噪比测量。

Signal-to-noise measurements in magnitude images from NMR phased arrays.

作者信息

Constantinides C D, Atalar E, McVeigh E R

机构信息

Department of Biomedical Engineering, Johns-Hopkins University School of Medicine, Baltimore, Maryland, USA.

出版信息

Magn Reson Med. 1997 Nov;38(5):852-7. doi: 10.1002/mrm.1910380524.

Abstract

A method is proposed to estimate signal-to-noise ratio (SNR) values in phased array magnitude images, based on a region-of-interest (ROI) analysis. It is shown that the SNR can be found by correcting the measured signal intensity for the noise bias effects and by evaluating the noise variance as the mean square value of all the pixel intensities in a chosen background ROI, divided by twice the number of receivers used. Estimated SNR values are shown to vary spatially within a bound of 20% with respect to the true SNR values as a result of noise correlations between receivers.

摘要

提出了一种基于感兴趣区域(ROI)分析来估计相控阵幅度图像中信噪比(SNR)值的方法。结果表明,可以通过校正测量信号强度的噪声偏差效应,并将噪声方差评估为所选背景ROI中所有像素强度的均方值,再除以所用接收器数量的两倍来找到SNR。由于接收器之间的噪声相关性,估计的SNR值在空间上相对于真实SNR值在20%的范围内变化。

相似文献

1
Signal-to-noise measurements in magnitude images from NMR phased arrays.
Magn Reson Med. 1997 Nov;38(5):852-7. doi: 10.1002/mrm.1910380524.
2
Adaptively optimized combination (AOC) of magnetic resonance spectroscopy data from phased array coils.
Magn Reson Med. 2016 Jun;75(6):2235-44. doi: 10.1002/mrm.25786. Epub 2015 Jul 20.
3
Increasing the signal-to-noise ratio by using vertically stacked phased array coils for low-field magnetic resonance imaging.
IEEE Trans Inf Technol Biomed. 2012 Nov;16(6):1150-6. doi: 10.1109/TITB.2012.2197633. Epub 2012 May 3.
4
Effects of proximity and noise level of phased array coil elements on overall signal-to-noise in parallel MR spectroscopy.
Magn Reson Imaging. 2018 Apr;47:125-130. doi: 10.1016/j.mri.2017.12.001. Epub 2017 Dec 5.
5
Noise correlations and SNR in phased-array MRS.
NMR Biomed. 2010 Jan;23(1):66-73. doi: 10.1002/nbm.1429.
6
The NMR phased array.
Magn Reson Med. 1990 Nov;16(2):192-225. doi: 10.1002/mrm.1910160203.
7
Image reconstruction in SNR units: a general method for SNR measurement.
Magn Reson Med. 2005 Dec;54(6):1439-47. doi: 10.1002/mrm.20713.
8
An NMR phased array for human cardiac 31P spectroscopy.
Magn Reson Med. 1992 Nov;28(1):54-64. doi: 10.1002/mrm.1910280106.
9
Intensity correction of phased-array surface coil images.
Magn Reson Med. 1996 Apr;35(4):585-90. doi: 10.1002/mrm.1910350419.
10
An unbiased signal-to-noise ratio measure for magnetic resonance images.
Med Phys. 1993 Jul-Aug;20(4):1077-8. doi: 10.1118/1.597004.

引用本文的文献

1
Rician Likelihood Loss for Quantitative MRI With Self-Supervised Deep Learning.
NMR Biomed. 2025 Oct;38(10):e70136. doi: 10.1002/nbm.70136.
2
ACID: A comprehensive toolbox for image processing and modeling of brain, spinal cord, and ex vivo diffusion MRI data.
Imaging Neurosci (Camb). 2024 Sep 13;2. doi: 10.1162/imag_a_00288. eCollection 2024.
3
Efficient PCA denoising of spatially correlated redundant MRI data.
Imaging Neurosci (Camb). 2023 Dec 18;1. doi: 10.1162/imag_a_00049. eCollection 2023.
4
Neural networks with personalized training for improved MOLLI T mapping.
BMC Med Imaging. 2025 Jul 1;25(1):245. doi: 10.1186/s12880-025-01769-z.
8
Feasibility and optimization ofF MRI on a clinical 3T with a large field-of-view torso coil.
Phys Med Biol. 2024 Jun 4;69(12):125002. doi: 10.1088/1361-6560/ad4d50.
9
The "hidden noise" problem in MR image reconstruction.
Magn Reson Med. 2024 Sep;92(3):982-996. doi: 10.1002/mrm.30100. Epub 2024 Apr 4.

本文引用的文献

1
The Rician distribution of noisy MRI data.
Magn Reson Med. 1995 Dec;34(6):910-4. doi: 10.1002/mrm.1910340618.
2
An unbiased signal-to-noise ratio measure for magnetic resonance images.
Med Phys. 1993 Jul-Aug;20(4):1077-8. doi: 10.1118/1.597004.
3
The use of power images to perform quantitative analysis on low SNR MR images.
Magn Reson Imaging. 1993;11(7):1051-6. doi: 10.1016/0730-725x(93)90225-3.
4
A phased array coil for human cardiac imaging.
Magn Reson Med. 1995 Jul;34(1):92-8. doi: 10.1002/mrm.1910340114.
5
Calculation of the signal-to-noise ratio for simple surface coils and arrays of coils.
IEEE Trans Biomed Eng. 1995 Sep;42(9):908-17. doi: 10.1109/10.412657.
6
A signal-to-noise calibration procedure for NMR imaging systems.
Med Phys. 1984 Mar-Apr;11(2):180-5. doi: 10.1118/1.595484.
7
Noise and filtration in magnetic resonance imaging.
Med Phys. 1985 Sep-Oct;12(5):586-91. doi: 10.1118/1.595679.
8
Measurement of signal intensities in the presence of noise in MR images.
Med Phys. 1985 Mar-Apr;12(2):232-3. doi: 10.1118/1.595711.
10
The NMR phased array.
Magn Reson Med. 1990 Nov;16(2):192-225. doi: 10.1002/mrm.1910160203.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验